Chapter 4:
Replacement Procedures
21
Control Panel Subassembly Replacement
Required Parts
One of the following:
• P08856 Service Assembly, Control Panel M-Turbo, English
• P08878 Service Assembly, Control Panel M-Turbo, French
• P08879 Service Assembly, Control Panel M-Turbo, German
• P08880 Service Assembly, Control Panel M-Turbo, Italian
• P08881 Service Assembly, Control Panel M-Turbo, Spanish
• P08882 Service Assembly, Control Panel M-Turbo, Portuguese
Required Tools
• #1 Phillips screwdriver
• Torque screwdriver, 2.0–10.0 inch pounds (0.23–1.1 newton meter)
• An anti-static mat
• A wrist grounding strap
Control Panel Removal
Control Panel Replacement
Caution:
Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at first.
Initial ESD symptoms may be slightly degraded performance or image quality.
Control Panel
Removal
1
Remove the two screws from the rear of the system per
2
Remove the two screws from the bottom of the system per
3
Turn the system over, fully open the display, and lift off the Control Panel per
.
Control Panel
Replacement
1
Place the new control panel in place.
2
Install the four screws removed in
“Control Panel Removal” on page 21
. Torque the screws
to 5.5 inch pounds.
Содержание M-Turbo
Страница 1: ...M Turbo Ultrasound System Service Manual TM...
Страница 6: ...2 Chapter 1 Introduction...
Страница 20: ...16 Chapter 3 Troubleshooting...
Страница 36: ...32 Chapter 4 Replacement Procedures...
Страница 44: ...40 Chapter 5 Performance Testing...
Страница 48: ...44 Appendix A Replacement Parts List Figure A 1 Power Supply P08850 Figure A 2 Speaker Assembly P03872...
Страница 52: ...48 Appendix B Service Event Report Service Event Report Form...
Страница 56: ...52 Index...
Страница 57: ......
Страница 58: ...P08144 01 P08144 01...