14
•
results:
result correct,
result below measuring range (under-range),
no result.
frequency weighting change: A / C (only if the measurement is stopped),
hold down 3 seconds – language change: POLISH / ENGLISH,
time weighting change: SLOW / FAST (only if the measurement is stopped),
START / STOP of measurement,
current results screen change,
hold down 1.5 second – clear current results (only if the measurement is stopped),
transition to measurement memory menu (only if the measurement is stopped),
hold down 1.5 second – transition to time history memory menu.
transition to date and time setup
hold down 1.5 second – transition to calibration menu (only if the measurement is
stopped).
WARNING! Changing frequency weighting or time weighting clear current measure results.
WARNING! Changing the language displays instrument data plate. After pressing any key,
results screen is shown back.
3.3. Date and time
You can use this menu to check the date and time currently set and change it. When you enter
edit mode, set the hour, minute, second, day, month and year respectively, passing to the next
item with the ENTER key. Pressing ENTER when the year is edited, causes saving date and time
and to leave edit mode.
WARNING! It is not possible to set invalid date (e.g. 31-02-2017). In that case, press the
Cancel button and set the date again.
Fig. 3.7. Date and time.
Fig. 3.8. Date and time - edition.
3.4. Measures memory
Using measures memory consists of three actions: save measure, load measure and delete
measure. It is also possible to delete all measures, but this requires additional confirmation.