Specifications
Section
6
6 - 4
a depth of 50 mm; at an acquisition of 128 rays,
the depth is 60 mm. Minimum is 18 mm.
Acoustic Output Reporting Tables
Typical measured results are below. There is a
considerable variation in transducers, and any
individual transducer may be quite different.
Table 6-1
A-Scan Acoustic Output Table
MI
I
SPTA.3
[mW/cm
2
]
I
SPPA.3
[W/cm
2
]
Acoustic Output – Global Maximum Value
0.085
0.0068
3.31
P
r.3
[MPa]
0.300
W
0
[mW]
0.00134
0.00134
f
c
[MHz]
12.29 12.29 12.29
z
sp
[cm]
1.7 1.7 1.7
Beam Dimension – x
-6
[cm]
0.383
0.383
Beam Dimension – y
-6
[cm]
0.310
0.310
PD
[
µ
s]
0.107
0.107
PRF
[Hz]
19.2
19.2
EBD Az.
[cm]
0.47
Associated
Acoustic
Parameters
EBD Ele.
[cm]
0.47
Table 6-2
B-Scan Acoustic Output Table
MI
I
SPTA.3
[mW/cm
2
]
I
SPPA.3
[W/cm
2
]
Acoustic Output – Global Maximum Value
0.154
0.881
11.45
P
r.3
[MPa]
0.475
W
0
[mW]
0.133
0.133
f
c
[MHz]
9.47 9.47 9.47
z
sp
[cm]
1.5
1.5
Beam Dimension – x
-6
[cm]
0.088
Beam Dimension – y
-6
[cm]
0.090
PD
[
µ
s]
0.135
0.135
PRF
[Hz]
3800
3800
EBD Az.
[cm]
0.60
Associated
Acoustic
Parameters
EBD Ele.
[cm]
0.60