![Sonomed E-Z Scan AB5500 Plus Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/sonomed/e-z-scan-ab5500-plus/e-z-scan-ab5500-plus_operators-manual_1322322036.webp)
Maintenance and Service
Section 5
5 - 4
WARNING
To avoid personal injury, do not open system –
there are no internal user-serviceable parts.
1. Check all cables and connections.
2. Check troubleshooting chart (Table 5-1) and
take any suggested corrective action.
Table 5-1
Troubleshooting Chart
Symptom
Corrective Action
Ensure AC adapter
connected to system and
proper power source
LCD screen does not
illuminate when system
powered on
Ensure power switch is in
the ON position
Verify that the
appropriate capture mode
has been selected (i.e.
cataract, dense cataract,
etc.)
Unable to freeze A-
scans in the automatic
capture mode
( only)
Verify Gain level is
adjusted properly
No refractive data is
shown in the IOL table
( only)
Verify that all necessary
IOL data have been
entered (i.e. k-readings,
formula constants, etc.)
Inconsistent axial length
measurements
obtained.
( only)
Check anterior chamber
depth measurements – if
inconsistent, may reflect
corneal compression
occurring during
applanation
Orbital pattern beyond
the retina does not
gradually decline
( only)
Ensure the probe is
aligned along the visual
axis
Verify B-probe is
connected properly
No image on B-scan
Display
Verify Footpedal is
connected properly
3. If problem cannot be solved, please contact
Sonomed service department for further
help. Prior to contacting Sonomed, please
gather as much information as possible
concerning the specific problem, and have
the system and probes nearby. This will
greatly aid the service representative in
determining the cause of the problem.
If it is deemed necessary to return the system
and/or probes to Sonomed for service, a
return authorization will be issued. Please
ensure that any materials returned to
Sonomed are adequately packaged to avoid
any damage during shipment.
ELECTROMAGNETIC INTERFERENCE
At times during normal use, the operator may
experience periods of interruption caused by
interference from outside sources such as
electrostatic discharge, power generators or
other office equipment.
Should any outside interference cause the
system to “lock-up” or begin to operate in a
manner inconsistent with the manufacturers
instructions, simply turning the unit off for a few
seconds will usually resolve the problem.
In some cases a “hard-reset” must be performed
by the operator. This will cause all stored data
to be replaced with the factory default settings,
and should therefore only be used when
absolutely necessary. For instructions on
performing a system reset, the operator should
contact the Sonomed service department at 800-
227-1285.
5.3
TROUBLESHOOTING
!