13
1.
Turn the liquid delivery system on.
2.
To energize the ultrasonics, press the
START
key. To de-energize the
ultrasonics, press the
STOP
key. To increase or decrease the amplitude in
small increments when the ultrasonics is on, depress the
ENTER/REVIEW
key repeatedly until the screen displays AMPLITUDE CONTROL, then
depress the
▲
or
▼
key, as required.
3.
Empirically determine the best flow rate / amplitude combination.
4.
If a footswitch or remote actuation device is used, depress the
switch to turn the ultrasonics off, plug in the footswitch or remote actuation
device into the footswitch jack, and depress the switch again to turn
the ultrasonics back on. Depressing the footswitch, or closing the contact on
the remote actuation device connected to the footswitch jack, will energize the
ultrasonics.
NOTE
For the majority of atomizing applications, low amplitude will atomize the
sample better than high amplitude.
NOTE
To clear an erroneous entry, press the
CLEAR
key.
NOTE
The
START
key and footswitch are mutually exclusive. If the process is initiated
by the
START
key, the footswitch becomes inoperative. If the process is initiated
by the footswitch, the
STOP
key becomes inoperative.
O
|