Sonics 830-00905 Скачать руководство пользователя страница 9

 

 

 

R

EV

.

  

12-21

 

3.3

 

A

IR 

C

ONSIDERATIONS

 

The  air  inlet  and  outlet  are  located  on  the  left  and  right  sides 
respectively.    Restricting  airflow  into or out of the unit will  impair 
performance.  At least 3 inches of clearance is required on each side 
to ensure adequate airflow. 
 

 

3.4

C

OOLANT 

F

ILL 

(

FOR THE 

830-00905

 CHILLER ONLY

Recommended Coolants:   

Sonics recommends DI water only.  For cup horn or microplate horn 
cooling,  DI  water  only  is  required.    Please  contact  a  Sonics 
representative  before  using  any  other  coolants  as  they  may  not  be 
compatible with the cup horn or microplate horn accessories. 

 

 

 

3.4

C

OOLANT 

F

ILL 

(

FOR THE 

830-00906

 CHILLER ONLY

The coolant fill cap is located at the top rear of the unit.  Twist off 
the cap counter-clockwise to open.  Fill reservoir to just below the 
bottom of its neck with coolant. Replace cap and close before 
operating chiller. 
 
If using the #830-00906 Chiller with a coolant other than water 
discontinue using the filter on the back of the unit.  Connect the 
Coolant Return tubing from the flow cell water jacket directly to the 
chiller Coolant Return port bypassing the filter. 

 

Recommended Coolants: 

 

DI water, ethylene glycol/water mixtures or Koolance, a pre-mixed 
27%  propylene  glycol/water  based  coolant  containing  an  algaecide 
and corrosion inhibitors, are acceptable coolants.  

Koolance comes in several colors, the colorless or blue versions in 
700 ml  bottles, part number:  LIQ-702CL-B (clear) or LIQ-702B-B 
(blue).    These  versions  are  recommended  as  the  dyes  in  the  other 
colored versions can form small particulates when not well mixed. 

 

Contact Koolance for details: 

Koolance USA 
2840 West Valley Highway North 
Auburn, WA 98001 
(253) 893-7551 
 

Note that algae/bacteria growth can occur when water is used 
alone as a coolant. Use AquaClear from Belart 

https://www.belart.com/aqua-clear-water-conditioner-
cleanware.html

 or similar that does not damage acrylic, to treat 

and keep water in the system clean.

 

Use only recommended 

coolants 

CAUTION 

WARNING 

Read the Coolant MSDS 

Prior to filling the chiller 

Use only recommended 

coolants 

CAUTION 

Содержание 830-00905

Страница 1: ...Recirculating Chiller Model 830 00905 830 00906 Operation Manual Sonics Materials Inc 53 Church Hill Road Newtown CT 06470 USA Phone 203 270 4600 info sonics com www sonics com Rev 12 21...

Страница 2: ..._________________________ 8 START UP__________________________________________________________________________ 8 SECTION 5 _________________________________________________________________________ 8 O...

Страница 3: ...ets the provisions of the directives 2004 108 EC EMC Directive 2006 95 EC Low Voltage Directive EN 61326 1 2006 Emissions and Immunity EN 61000 3 2 2006 Harmonics Emissions EN 61000 3 3 2008 Voltage F...

Страница 4: ...ng point Only water is recommended when using a 830 00905 with an ultrasonic processor and cup horn microplate horn accessory When using a 830 00906 always use only proper coolants as specified in man...

Страница 5: ...s The system provides 1 liter per minute of constant temperature coolant with PID control for both cooling and heating With fewer moving parts the system is highly reliable and energy efficient The ch...

Страница 6: ...00906 DI Water Koolance 27 propylene glycol water mix or 27 50 ethylene glycol water mix Process Fluid Fittings 1 4 quick connect type connectors Colder brand Pumps Diaphragm Wetted Materials Aluminu...

Страница 7: ...830 00905 830 00906 Side Figure 3C 830 00905 830 00906 Back Filter Wrench used to tighten or remove filter housing Process Fluid Connectors On Off Switch Signal Connector 9 Pin D sub Display Screen Co...

Страница 8: ...20mm GDB quick acting glass meets IEC 127 2 Replacement Fuse Sonics 20 22332 10 Allied Electronics 70149445 Optional Alarms Alarm signals are TTL signals normally high 4 VDC located on the 9 pin d sub...

Страница 9: ...nue using the filter on the back of the unit Connect the Coolant Return tubing from the flow cell water jacket directly to the chiller Coolant Return port bypassing the filter Recommended Coolants DI...

Страница 10: ...LEVEL LOW add more coolant to the reservoir until the display changes to read the coolant temperature Important Notes Do not run the chiller dry for more than a few seconds SECTION 5 OPERATION _______...

Страница 11: ...to the set point temperature To change the set point temperatures just press the UP or DOWN keys again to change SETTEMP 1 to the new set point followed by ENTER and then START STOP The Operating Mode...

Страница 12: ...lly TTL low 0 5 VDC system failure has occurred A list of system failures causing the system alarm to change to TTL low can be found in Section 6 In the event of a system failure the alarm type will b...

Страница 13: ...for a RMA number to return the unit for fan replacement Pump Fail The liquid heat exchanger plate temperature is either too hot or too cold indicating pump failure a blockage in the external plumbing...

Страница 14: ...larms are present poor temperature control indicates the heat load is too great for the chiller the TE cooling heating engine is not receiving power the PID constants have been corrupted or the chille...

Страница 15: ...non shedding wipe dipped in isopropyl alcohol SECTION 9 TECHNICAL SUPPORT _____________________________________________ Customer service is our highest priority Please contact us immediately for tech...

Страница 16: ...ich invalidate the warranty include but are not limited to Removing the serial number label Any disassembly partial or complete of the product Changing any components of the product Subjecting the pro...

Отзывы: