The  chiller  has  multiple  system  alarms  that  when  triggered  will 
show  on the display.  When an alarm is displayed the system will 
not attempt to heat or cool the coolant.   

 

Alarms:

 

RTD  Open:  The  temperature  sensor  has  failed  or  its  connector  has 
come loose.   

Turn off the chiller and disconnect the AC power cord.

  

Contact  Sonics  for  a  replacement  RTD,  or  for  a  RMA  number  to 
return the unit for RTD replacement.

 

Fan Fail: Fan is supplying insufficient air to cool the thermoelectric 
heat exchanger.  

Either the fan has failed or the airflow into or out 

of  the  system  is  blocked.  Check  that  the  side  air  inlet  and  outlet 
gratings are not blocked.  The chiller requires at least 3 inches of 
clearance around these gratings.  If airflow is not blocked, contact 
Sonics  for  a  replacement  fan,  or  for  a  RMA number  to  return  the 
unit for fan replacement.

 

Pump Fail: The liquid heat exchanger plate temperature is either too 
hot or too cold, indicating pump failure, a blockage in the external 
plumbing lines or operation outside the normal 5°C to 50°C coolant 
temperature (without –LT or –HT options). 

Turn off the chiller and 

disconnect  the  AC  power  cord.    Verify that  no  kinks  or  blockages 
exist  in  plumbing  line,  both  outside  and  inside  the  chiller.    If  no 
coolant  flow  blockages  exist,  contact  Sonics  for  a  replacement 
pump,  or  for  a  RMA  number  to  return  the  unit  for  pump 
replacement. 

Tank  Level  Low(#830-00906  Only):  Liquid  reservoir  level  is  too 
low. 

Unless  filling  for  the  first  time,  check  all  outside  plumbing 

lines for leaks.  Once all leaks are sealed, remove the cap and add 
more coolant until the alarm disappears.

 

 

Electrical Shock Hazard:  

Always unplug the unit 

before removing the cover.

 

WARNING 

Do not attempt to service or 

repair the unit beyond the 

troubleshooting checks 

described in this section 

without first contacting  

Solid State Cooling Systems

 

WARNING 

Содержание 830-00905

Страница 1: ...Recirculating Chiller Model 830 00905 830 00906 Operation Manual Sonics Materials Inc 53 Church Hill Road Newtown CT 06470 USA Phone 203 270 4600 info sonics com www sonics com Rev 12 21...

Страница 2: ..._________________________ 8 START UP__________________________________________________________________________ 8 SECTION 5 _________________________________________________________________________ 8 O...

Страница 3: ...ets the provisions of the directives 2004 108 EC EMC Directive 2006 95 EC Low Voltage Directive EN 61326 1 2006 Emissions and Immunity EN 61000 3 2 2006 Harmonics Emissions EN 61000 3 3 2008 Voltage F...

Страница 4: ...ng point Only water is recommended when using a 830 00905 with an ultrasonic processor and cup horn microplate horn accessory When using a 830 00906 always use only proper coolants as specified in man...

Страница 5: ...s The system provides 1 liter per minute of constant temperature coolant with PID control for both cooling and heating With fewer moving parts the system is highly reliable and energy efficient The ch...

Страница 6: ...00906 DI Water Koolance 27 propylene glycol water mix or 27 50 ethylene glycol water mix Process Fluid Fittings 1 4 quick connect type connectors Colder brand Pumps Diaphragm Wetted Materials Aluminu...

Страница 7: ...830 00905 830 00906 Side Figure 3C 830 00905 830 00906 Back Filter Wrench used to tighten or remove filter housing Process Fluid Connectors On Off Switch Signal Connector 9 Pin D sub Display Screen Co...

Страница 8: ...20mm GDB quick acting glass meets IEC 127 2 Replacement Fuse Sonics 20 22332 10 Allied Electronics 70149445 Optional Alarms Alarm signals are TTL signals normally high 4 VDC located on the 9 pin d sub...

Страница 9: ...nue using the filter on the back of the unit Connect the Coolant Return tubing from the flow cell water jacket directly to the chiller Coolant Return port bypassing the filter Recommended Coolants DI...

Страница 10: ...LEVEL LOW add more coolant to the reservoir until the display changes to read the coolant temperature Important Notes Do not run the chiller dry for more than a few seconds SECTION 5 OPERATION _______...

Страница 11: ...to the set point temperature To change the set point temperatures just press the UP or DOWN keys again to change SETTEMP 1 to the new set point followed by ENTER and then START STOP The Operating Mode...

Страница 12: ...lly TTL low 0 5 VDC system failure has occurred A list of system failures causing the system alarm to change to TTL low can be found in Section 6 In the event of a system failure the alarm type will b...

Страница 13: ...for a RMA number to return the unit for fan replacement Pump Fail The liquid heat exchanger plate temperature is either too hot or too cold indicating pump failure a blockage in the external plumbing...

Страница 14: ...larms are present poor temperature control indicates the heat load is too great for the chiller the TE cooling heating engine is not receiving power the PID constants have been corrupted or the chille...

Страница 15: ...non shedding wipe dipped in isopropyl alcohol SECTION 9 TECHNICAL SUPPORT _____________________________________________ Customer service is our highest priority Please contact us immediately for tech...

Страница 16: ...ich invalidate the warranty include but are not limited to Removing the serial number label Any disassembly partial or complete of the product Changing any components of the product Subjecting the pro...

Отзывы: