Recirculating Chiller
Model #830-00905 / 830-00906
Operation Manual
Sonics & Materials, Inc.
53 Church Hill Road
Newtown, CT 06470 USA
Phone: 203.270.4600
[email protected]
www.sonics.com
Rev. 12-21
Страница 1: ...Recirculating Chiller Model 830 00905 830 00906 Operation Manual Sonics Materials Inc 53 Church Hill Road Newtown CT 06470 USA Phone 203 270 4600 info sonics com www sonics com Rev 12 21...
Страница 2: ..._________________________ 8 START UP__________________________________________________________________________ 8 SECTION 5 _________________________________________________________________________ 8 O...
Страница 3: ...ets the provisions of the directives 2004 108 EC EMC Directive 2006 95 EC Low Voltage Directive EN 61326 1 2006 Emissions and Immunity EN 61000 3 2 2006 Harmonics Emissions EN 61000 3 3 2008 Voltage F...
Страница 4: ...ng point Only water is recommended when using a 830 00905 with an ultrasonic processor and cup horn microplate horn accessory When using a 830 00906 always use only proper coolants as specified in man...
Страница 5: ...s The system provides 1 liter per minute of constant temperature coolant with PID control for both cooling and heating With fewer moving parts the system is highly reliable and energy efficient The ch...
Страница 6: ...00906 DI Water Koolance 27 propylene glycol water mix or 27 50 ethylene glycol water mix Process Fluid Fittings 1 4 quick connect type connectors Colder brand Pumps Diaphragm Wetted Materials Aluminu...
Страница 7: ...830 00905 830 00906 Side Figure 3C 830 00905 830 00906 Back Filter Wrench used to tighten or remove filter housing Process Fluid Connectors On Off Switch Signal Connector 9 Pin D sub Display Screen Co...
Страница 8: ...20mm GDB quick acting glass meets IEC 127 2 Replacement Fuse Sonics 20 22332 10 Allied Electronics 70149445 Optional Alarms Alarm signals are TTL signals normally high 4 VDC located on the 9 pin d sub...
Страница 9: ...nue using the filter on the back of the unit Connect the Coolant Return tubing from the flow cell water jacket directly to the chiller Coolant Return port bypassing the filter Recommended Coolants DI...
Страница 10: ...LEVEL LOW add more coolant to the reservoir until the display changes to read the coolant temperature Important Notes Do not run the chiller dry for more than a few seconds SECTION 5 OPERATION _______...
Страница 11: ...to the set point temperature To change the set point temperatures just press the UP or DOWN keys again to change SETTEMP 1 to the new set point followed by ENTER and then START STOP The Operating Mode...
Страница 12: ...lly TTL low 0 5 VDC system failure has occurred A list of system failures causing the system alarm to change to TTL low can be found in Section 6 In the event of a system failure the alarm type will b...
Страница 13: ...for a RMA number to return the unit for fan replacement Pump Fail The liquid heat exchanger plate temperature is either too hot or too cold indicating pump failure a blockage in the external plumbing...
Страница 14: ...larms are present poor temperature control indicates the heat load is too great for the chiller the TE cooling heating engine is not receiving power the PID constants have been corrupted or the chille...
Страница 15: ...non shedding wipe dipped in isopropyl alcohol SECTION 9 TECHNICAL SUPPORT _____________________________________________ Customer service is our highest priority Please contact us immediately for tech...
Страница 16: ...ich invalidate the warranty include but are not limited to Removing the serial number label Any disassembly partial or complete of the product Changing any components of the product Subjecting the pro...