CHARGER miniRITE T R · USER GUIDE
Symptom
Possible causes
Solutions
The charger LED
remains turned off
when the charger is
connected to power
The charger is not plugged in
Verify that the power supply is connected correctly.
Make sure that the power source complies with specifica-
tions indicated on page 15.
The charger LED shows
long red blinks
The hearing aid or charger is too
warm
Move the charger and hearing aid to a location with a
temperature b5 °C to +40 °C (+41 °F to +104 °F)
The charger LED
repeatedly shows
3 short red blinks
The charger is too cold
Move the charger and hearing aid to a location with a
temperature b5 °C to +40 °C (+41 °F to +104 °F)
System error
Contact your hearing care professional
Hearing aid LED
remains off when the
hearing aid is placed in
the charger
The charger is not plugged in
Verify that the charger power supply is connected correctly
The hearing aid or charger is either
too warm or too cold
Move the charger and hearing aid to a location with a
temperature b5 °C to +40 °C (+41 °F to +104 °F)
Charging is incomplete. The room
temperature e35 °C
(+95 °F), which prolongs the
charging time. The charger has
stopped charging to protect the
battery.
Reinsert the hearing aid into the charger. This will complete
the charging within approximately 15 minutes.
The hearing aid is not correctly
inserted into the charger
Check the charger ports for debris
The hearing aid LED
blinks red when the
hearing aid is placed in
the charger
System error
Contact your hearing care professional