CMM-60
– USER MANUAL
51
To set other reference value in
dBm
mode:
in the measurement menu, select
REF
,
use
arrows
▲▼
to c=select desired reference value: 4, 8, 16, 25, 32, 50, 75, 600 or 1000
Ω,
select
OK.
6.1.3 Low pass filter
The meter is equipped with a low pass filter of alternating voltage with the threshold frequency of 1
kHz. This function blocks any voltage with a frequency exceeding 1 kHz. Voltages with a frequency
below the threshold are taken into account with reduced accuracy. The filter is useful in measuring the
signals of multiple sinusoidal sub-signals.
In voltage measurement mode for
range of 1000 V
:
select
MENU,
use
arrows
and
F1 F2 F3 F4
buttons to select
.
6.2 Voltage measurement up to 500 mV
To perform voltage measurement:
set the rotary switch at
mV
Temp
,
select
MENU
and use
arrows
and
F1 F2 F3 F4
buttons to set the measurement of:
direct current voltage
mVDC
,
alternating current voltage
mVAC
(default setting),
connect
black
test lead to
COM
terminal and
red
test lead to
V
Ω
CAP Hz% Temp
terminal,
contact the tips of test probes to the points of measurement; red probe should be applied to the
point of higher potential,
read the measurement result on the display,
after completing the measurements, remove test leads from the terminals of the meter.
6.3 Temperature Measurement
To perform the measurement:
set the rotary switch at
mV
Temp
,
select
MENU,
use
arrows
and
F1 F2 F3 F4
buttons to set the temperature measurement
Temp
and unit
– Cel-
sius
C
or Fahrenheit
F
,
place the adapter of the temperature probe
in
COM
terminal (black leg) and
V
Ω
CAP
Hz% Temp
(red leg):
place the temperature probe
in the
adapter,
as shown in the figure:
thin pin of the probe (marked as
+
) fits to terminal
+
;
thick pin of the probe (marked as
K
) fits to terminal
–
;
reversed connection of the probe is mechanically
impossible
,
contact the head of the temperature probe to the device under test. Maintain the contact of the
probe head with the part of the device under test, until the reading stabilizes.
read the measurement result on the display,
after completing the measurements, disconnect the probe from the meter.