CMM-40
– USER MANUAL
59
Caution:
The temperature probe is fitted with a type K mini connector.
A mini connector to banana connector adaptor is supplied
for connection to the input banana jacks.
5.11 Frequency or % duty cycle measurements
In order to realise frequency or % duty cycle measurement it is nec-
essary to realise the following actions:
Set the rotary function switch to the green
Hz/%
position,
Insert the black lead banana plug into the negative
COM
jack
and the red test lead banana plug into the positive
Hz
jack,
Touch the test probe tips to the circuit under test
,
Read the frequency on the display,
Press the
MODE
button to indicate “
%
”,
Read the % duty cycle in the display,
Having done the measurement disconnect the test leads from
the meter.
5.12 % 4
– 20 mA measurements
In order to realise measurement it is necessary to realise the follow-
ing actions:
Set up and connect as described for DC mA measurements,
Set the rotary function switch to the
4-20mA%
position,
The meter will display loop current as a % with 0mA=-25%,
4mA=0%, 20mA=100%, and 24mA=125%.Special functions
6
Special functions
6.1 Autoranging/manual range selection
When the meter is first turned on, it automatically goes into au-
toranging. This automatically selects the best range for the meas-
urements being made and is generally the best mode for most
measurements. For measurement situations requiring that a range
be manually selected, perform the following:
Содержание CMM-40
Страница 1: ...INSTRUKCJA OBS UGI 1 USER MANUAL 39 MANUAL DE USO 73 CMM 40 v1 6 19 10 2018...
Страница 2: ......
Страница 3: ...1 INSTRUKCJA OBS UGI MULTIMETR PRZEMYS OWY CMM 40 Wersja 1 6 19 10 2018...
Страница 41: ...39 USER MANUAL INDUSTRIAL MULTIMETER CMM 40 Version 1 6 19 10 2018...
Страница 75: ...73 MANUAL DE USO MULT METRO INDUSTRIAL CMM 40 Versi n 1 6 19 10 2018...
Страница 111: ......
Страница 112: ......