-104-
9)
% Snore time
If the percentage of snoring, during an apnoea phase, exceeds this value, a hypopnoea will be
detected.
10)
Body position correlation [s]
With this function the correlation of a flow event is checked with body position change. The adjustable
delay gives the maximum time between a Flow event and a body position change.
11)
Intermediate Wake
Activating this function will count an Apnoea/Hypopnoea event as sleep during the first wake stage
after a respiratory event. The event however must start in a sleep stage.
7.3.4.8 Snore Analysis
fig. 7-17: Snore Analysis
1)
Analysis Window [s]
Time period used by the analysis to determine the baseline.
2)
Ratio [%]
This ratio
shows the percentage increase by which the signal needs to exceed the baseline. All values
above this ratio are detected as snore.
3)
Min. Duration [ms]
Minimum duration of a snore event.
4)
Max. Duration [ms]
Maximum duration of a snore event.
Содержание SOMNOwatch plus
Страница 24: ...24 Body Temperature fig 6 8 Montage Temperature EDA fig 6 9 Montage EDA...
Страница 25: ...25 Sleep Profile fig 6 10 Montage Sleep Profile ECG Analysis fig 6 11 Montage ECG...
Страница 26: ...26 Training Rehab fig 6 12 Montage Training Rehab Flow Snore fig 6 13 Montage Flow Snore...
Страница 42: ...42 Body Temperature fig 6 22 Analysis Template Temperature EDA fig 6 23 Analysis Template EDA...
Страница 43: ...43 Sleep Profile fig 6 24 Analysis Template Sleep Profile...
Страница 45: ...45 Flow Snore fig 6 27 Analysis Template Flow Snore CPAP Control fig 6 28 Analysis Template CPAP...
Страница 63: ...63 Choose the item Activity profil within the Select Report window...
Страница 72: ...72 recording on both legs...
Страница 162: ...162 10 9 Notes...