1.7
ECT CABLE JACK
The ECT cable jack is a keyed round 7 pin plastic lock connector. The ECT cable plug is
connected to this jack.
1.8
EEG/EKG CABLE JACK
The EEG/EKG cable jack is a keyed round 9 pin plastic lock connector. The EEG/EKG
cable plug is connected to this jack.
1.9
FLEXDIAL
The Flexdial enables the user to change all parameters by pressing and turning the dial and
following the attached flow diagram. The settings are remembered in the battery-backed
RAM. The Flexdial LED blinks whenever the user enters the Flexdial mode. To exit,
press the START/STOP button.
1.10 START/STOP BUTTON
START/STOP button is used to control the operation of the printer. Pressing the button
will start the printer; another press will stop it. It is also used to terminate the Flexdial
mode. When the Flexdial mode is exited with the START/STOP button the system will
printout all the parameters.
1.11 100 MM THERMAL PRINTER
The printer is used to print out all graphic and alphanumeric data. I uses 100 mm fanfold
paper. To open the paper storage press the tab above the printer roller.
1.12 PRESET AND USER SET LED
Whenever the system is in the factory preset mode the PRESET LED is on. If the user sets
any treatment parameter to a setting different from the preset the USER SET LED is lit.
1.13 SAFETY MONITOR ACTIVATED LED
A separate safety monitoring microprocessor is used to monitor the output. It reads the
energy level and if the level is exceeded by 20% it trips an audio alarm and sets off the
LED. It also disables the output. The only way to reset this is to turn off the system.