SOMATICS, LLC 

 

 

 

 

 

 

 

REV: 3.1   09/15/2015

 

 

5

 

 

1.  EXPLANATIONS OF OPERATING CONTROLS 
 
 

FRONT PANEL 

 

See fig. 1.1 

 

 

1.1 

POWER SWITCH 
The power switch is an alternate position switch. When activated it connects both the hot 
and the neutral into the power circuits.  

 
 
1.2 

IMPEDANCE SWITCH 
The impedance switch is a momentary switch. When depressed, it signals the processor to 
start the impedance test sequence. After a second, an impedance readout is displayed. The 
display is on for the duration of the switch activation. 

 
 
1.3 

IMPEDANCE/TIME READOUT 
The impedance readout is an 8 digit alphanumeric LED (Light Emitting Display) which 
indicates the impedance of the stimulus electrodes when the impedance test switch is 
depressed. If the impedance is over 3000 Ohms the display blinks. After the treatment is 
delivered the same 8 digit LED is used to display time elapsed since the end of treatment.  
It is also used during the Flexdial setup mode to alter parameters. 

 
1.4 

PERCENT ENERGY READOUT 
The percent energy readout is a 4 digit alphanumeric LED (Light Emitting Display) which 
indicates the setting of the energy dial. It is also used to indicate the treatment program 
that is being used. 

 
1.5 

DOSAGE ADJUSTMENT DIAL 
The dosage adjustment dial sets the number of pulses in the stimulus pulse train. It's a 
continuously rotating optical encoder. In the default program setting of LOW 0.5 the 
lowest setting gives 28 pulses. Each increase in the setting of the dial adds 28 more pulses 
to the stimulus. The maximum setting is 560 pulses. The dial is calibrated in PERCENT 
ENERGY of its maximum for ease of use. After each change there is 1 second display of 
the millicoulomb value before the display returns to the percent display. If the dial is 
pressed the display will indicate the treatment program that is being used. 

 
1.6 

TREAT SWITCH 
The TREAT switch is a momentary switch which controls the delivery of the stimulus. It 
has a special protective cover which prevents accidental activation. As an additional safety 
feature there is a one second warning tone before the stimulus begins whenever the 
TREAT button is depressed. The yellow indicator light inside the button lights up only 
when the electrical stimulus is being delivered. The switch must be depressed for the 
duration of the stimulus. The delivery of the stimulus is indicated by the TREAT button 
light and a warbling sound. If the button is released, the stimulus will be immediately 
terminated. 

 

Содержание Thymatron System IV

Страница 1: ...U ARE QUALIFIED TO DO SO Document number SM TS4 REV 3 1 Date September 15 2015 Check our website at least once a year for updated User s Manual and Service Manual Downloads are available there Technical Support USA and Canada 1 800 642 6761 Telephone 1 847 234 6761 Fax 1 847 234 6763 E Mail sales thymatron com Web Site www thymatron com Please check for revisions and change of information at the r...

Страница 2: ...of Somatics LLC are covered by U S and foreign patents and or pending patents Thymatron Thymatron DG Thymatron DGx Thymatron System IV DualGraph Flexdial Crona are registered trademarks of Somatics LLC Class 1 Type BF Is the International symbol for Type BF and Defibrillator protected Is the International symbol for CAUTION Printed in U S A Specifications and price change privileges are reserved ...

Страница 3: ...AGE 1 EXPLANATION OF OPERATING CONTROLS 4 2 SAFETY CHECK 8 3 CALIBRATION CHECK 9 4 THEORY OF OPERATION 15 5 CALIBRATION AND FACTORY PRESETS 18 6 SPECIFICATIONS 21 7 CONNECTOR DEFINITIONS 22 8 PERCENT ENERGY TABLES 32 9 SCHEMATICS and ASSEMBLY 34 ...

Страница 4: ...SOMATICS LLC REV 3 1 09 15 2015 4 fig 1 1 fig 1 2 ...

Страница 5: ...ndicate the treatment program that is being used 1 5 DOSAGE ADJUSTMENT DIAL The dosage adjustment dial sets the number of pulses in the stimulus pulse train It s a continuously rotating optical encoder In the default program setting of LOW 0 5 the lowest setting gives 28 pulses Each increase in the setting of the dial adds 28 more pulses to the stimulus The maximum setting is 560 pulses The dial i...

Страница 6: ...g the button will start the printer another press will stop it It is also used to terminate the Flexdial mode When the Flexdial mode is exited with the START STOP button the system will printout all the parameters 1 11 100 MM THERMAL PRINTER The printer is used to print out all graphic and alphanumeric data I uses 100 mm fanfold paper To open the paper storage press the tab above the printer rolle...

Страница 7: ...1 23 LINE FUSES For 115V lines this instrument uses two 2 0 amp 250V T type 5x25 mm fuses For 230V lines this instrument uses two 1 0 amp 250V T type 5x25 mm fuses Do not replace with a different type of fuse 1 24 IEC POWER CONNECTOR Use the cable supplied and plug this instrument only into a well grounded three prong receptacle making sure you select the correct input voltage 1 25 INPUT VOLTAGE S...

Страница 8: ...ial leakage tester or by using a test box constructed so as to permit disconnecting ground neutral and reversing AC polarity as in figure 2 1 Test leakage under any combination of the switch settings All current readings should be less than the maximum reading in µamps RMS read according to chart below between any chassis or electrode point and the test box Leakage to Ground Source GROUND POWER PO...

Страница 9: ...0 ohms 10W resistor to the ECT banana plugs See Fig 3 1 fig 3 1 3 2 Connect Oscilloscope across the dummy load Be sure the scope can measure up to 500 volts If not construct a high impedance divider such as in fig 3 2 optional if available use 10x probe instead fig 3 2 CAUTION HIGH VOLTAGES ARE PRESENT DURING TESTING EXERCISE CAUTION TO PREVENT INJURY ...

Страница 10: ... to 0 5 msec div Set trigger for positive phase Turn power on Make sure the system is in default setting The program should be in LOW 0 5 Set PERCENT ENERGY dial to 100 and push TREAT button Observe wave shape as in figure 3 3 Peak voltage reading should be 180 Volts 7 5 which corresponds to 900 ma at 200 ohms For other loads wave shapes see fig 3 4 fig 3 3 ...

Страница 11: ...SOMATICS LLC REV 3 1 09 15 2015 11 fig3 4 ...

Страница 12: ...responds to 70 Hz Set PERCENT ENERGY dial to 50 and press TREAT button The reading should be 25 00 msec 5 which corresponds to 40 Hz Set PERCENT ENERGY dial to 20 and press TREAT button The reading should be 50 0 msec 5 which corresponds to 20 Hz 3 6 TEST STIMULUS DURATION There are two ways to test stimulus duration The more precise way uses an event counter many frequency counters are capable of...

Страница 13: ...0 sec div Set proper trigger Set PERCENT ENERGY dial to 100 and push the TREAT button The duration of the stimulus should be 8 seconds as in fig 3 6 fig 3 6 3 7 IMPEDANCE TEST Connect a variable resistor box to the ECT stimulus cable Set on zero ohms and push the impedance test button the readout should be 0_ 100_ Repeat for 1000_ and 2000_ At impedances higher than 3000 _ the display will read 30...

Страница 14: ... and note 3 db drop in amplitude fig 3 7 3 9 EEG AMPLIFIER NOISE TEST Connect three 5K_ resistors into the electrode input of the EEG amplifier according to fig 3 8 With the oscilloscope connected to the EEG output jack note that output noise is no more than 50 mv equivalent to 5µv at the input fig 3 8 3 10 EEG AMPLIFIER SOUND TEST With oscillator and oscilloscope connected connect dummy load to t...

Страница 15: ...d output measurement channels The digital data from and to the A D are isolated with a high speed optical isolators An isolated power supply provided 5 V from the 5V power The impedance measurement is derived by injecting 6 µamp current into the treatment electrodes and using an isolated differential amplifier to amplify the difference voltage which is proportional to the impedance of the electrod...

Страница 16: ...er supply provides power for the positive half of the ECT output drivers and for motor and thermal head of the printer c The 24V power supply provides power for the negative half of the ECT output drivers Each supply is individually protected by a fuse The ECT output section contains the constant current supply and two power drivers One driver controls positive pulses while the other driver contro...

Страница 17: ...CT CN502 EEG CN501 TREAT SW502 T101 POWER T102 OUTPUT IEC F1 F2 VOLUME PR101 SPEAKER POWER SW507 ANALOG BOARD CN202 CN201 CN203 CN402 CN205 PRINTER CN301 CN303 CN 304 CN401 CN602 SET ADJUST SW506 START STOP SW504 ALARM SW601 CN403 CN404 CN407 CN406 CN405 CN302 CN204 RS232 CN601 SW602 115 230V SWITCH CN106 CN206 DIAGNOSTIC fig 4 1 ...

Страница 18: ...F HIGH VOLTAGES ARE EXPOSED EXERCISE CAUTION TO AVOID INJURY 5 2 With the system turned off install a jumper on J3 test Turn on the system and note that the display shows CALIBR 5 3 EEG AMPLIFIER GAIN ADJUSTMENT Set oscillator to 2V p p and 10Hz Use a calibrated meter to adjust the oscillator to 707 Volts It may help to temporarily set the oscillator to 30 Hertz but remember to reset it back down ...

Страница 19: ...he low wattage load will be destroyed and possibly catch on fire if a treatment is initiated Press the IMPEDANCE button and hold until a value is displayed Make sure the reading is between 1350 and 1650 ohms If it is press the Flexdial button to calibrate the display should show 1500 Press the START STOP button to exit the impedance calibration and return to the main shell The display will read CA...

Страница 20: ...e one click and the display should show CONFIG Press the Flexdial button and rotate the Flexdial to select 50 or 60 Hz notch filter depending on the frequency of the power mains Press the Flexdial button again and rotate the Flexdial to select the date format either Day Month Year or Month Day Year Press the Flexdial button again and rotate the Flexdial to and select 24 hour or 12 hour format Pres...

Страница 21: ... dB points at 2 Hz and 25 Hz Notch filter 40 dB down at 60 Hz CMRR 80 dB EEG sound frequency 500 Hz modulated from channel 1 6 3 EMG channel 3 Amplification 100 20 mv cm 5 Bandwidth 3 dB points at 2 Hz and 50 Hz Notch filter 40 dB down at 60 Hz CMRR 80 dB 6 4 EKG channel 4 Amplification 100 20 mv cm 5 Bandwidth 3 dB points at 2 Hz and 50 Hz Notch filter 40 dB down at 50 or 60 Hz CMRR80 dB 6 5 Prin...

Страница 22: ... 9 isognd CN106 3pin SIP 1 OUTAO 2 3 OUTBO CN105 3pin SIP 1 OUTAI 2 3 OUTBI CN101 10pin SIP 1 speaker 2 speaker 3 volume high 4 volume center 5 volume low CN102 40 pin DIN 1 warn sound To Control 2 EEG sound enable EEGS 3 ALARM 4 GAIN 5 BD0 6 BD1 7 BD2 8 BD3 8 BD4 10 BD5 11 BD6 12 BD7 13 EADC 14 RD 15 WR 16 RST 17 GND 18 CLK 19 GND 20 ADERR ...

Страница 23: ... 2xEN 30 TREATSW 33 GND 34 GND 35 GND 36 GND 37 5V 38 5V 39 5V 40 5V 2 CONTROL BOARD CN201 40 pin DIN 1 warn sound To Analog 2 EEG sound enable EEGS 3 ALARM 4 GAIN 5 BD0 6 BD1 7 BD2 8 BD3 8 BD4 10 BD5 11 BD6 12 BD7 13 EADC 14 RD 15 WR 16 RST 17 GND 18 CLK 19 GND 20 ADERR 21 ADEF 22 ADDR 23 ADTR 24 ENERGY0 25 ENERGY1 ...

Страница 24: ...D 37 5V 38 5V 39 5V 40 5V CN202 40pin DIN 1 warn sound To Display 2 MENULED 3 GND 4 GAIN 5 BD0 6 BD1 7 BD2 8 BD3 8 BD4 10 BD5 11 BD6 12 BD7 13 EAND 14 RD 15 WR 16 RST 17 GND 18 CLK 19 GND 20 ENERGY0 21 ENERGY1 22 ENERGY2 23 ENERGY3 24 ENERGY4 25 2xEN 26 FLEX0 27 FLEX1 28 FLEX2 29 FLEX3 30 RLTREAT 31 TRIND 32 PLED ...

Страница 25: ...03 50pin HDDIN 1 PRRD To Printer 2 GND 3 BD0 4 GND 5 BD1 6 GND 7 BD2 8 GND 9 BD3 10 GND 11 BD4 12 GND 13 BD5 14 GND 15 BD6 16 GND 17 BD7 18 GND 19 BA0 20 GND 21 BA1 22 GND 23 BA2 24 GND 25 BA3 25 GND 27 BA4 28 GND 29 BA5 30 GND 31 BA6 32 GND 33 BA7 34 GND 35 BA8 36 GND 37 PRWR 38 GND 39 BUSY ...

Страница 26: ...SOMATICS LLC REV 3 1 09 15 2015 26 40 RST 41 ERROR 42 FEED 43 SOUT 44 SIN 45 INTRP 46 POST 47 BRDY 48 SUP 49 ABUFF 50 SUP CN204 10 pin DIN 1 To RS232 2 TXD 3 GND 4 RTS 5 6 RDX 7 DTR 8 9 CTS ...

Страница 27: ...D 11 OUTA 12 OUTB 3 POWER OUTPUT BOARD CN301 12pin SIP 1 5V To Control 2 5V 3 5V 4 5V 5 5V 6 GND 7 GND 8 GND 9 GND 10 GND 11 OUTA 12 OUTB CN302 6pin SIP 1 PGND To Printer 2 PGND 3 24V 4 24V 5 5V 6 LGND CN303 5pin SIP 1 T305 2 3 gnd 4 5 T304 CN304 10pin SIP 1 5Vac A 2 5Vac A 3 5Vac B 4 5Vac B 5 24Vac 6 24Vac 7 24Vac ...

Страница 28: ...LED 3 GND 4 GAIN 5 BD0 6 BD1 7 BD2 8 BD3 8 BD4 10 BD5 11 BD6 12 BD7 13 EAND 14 RD 15 WR 16 RST 17 GND 18 CLK 19 GND 20 ENERGY0 21 ENERGY1 22 ENERGY2 23 ENERGY3 24 ENERGY4 25 2xEN 26 FLEX0 27 FLEX1 28 FLEX2 29 FLEX3 30 TREATSW 31 TRIND 32 PLED 33 GND 34 GND 35 GND 36 GND 37 5V 38 5V 39 5V 40 5V CN402 5 pin SIP 1 TREATSW 2 GND ...

Страница 29: ... GND 3 CN406 6 pin SIP 1 GND 2 GND 3 ENERGYSW 4 ENOUTB 5 ENOUTA 6 5V CN405 6 pin SIP 1 GND 2 GND 3 SETSW 4 ADOUTB 5 ADOUTA 6 5V CN407 3 pin SIP 1 ALARMSW 2 GND 3 5 FRONT PANEL CN501 9pin Hyperion 1 ch1 EEG 2 ch1 3 ch2 4 ch2 5 ch3 6 ch3 7 ch4 8 ch4 9 isognd CN502 7pin Hyperion 1 ECT 2 OUTAO 3 OUTBO 4 5 RTREATA ...

Страница 30: ...Start Stop switch 1 STSTSW 2 GND SW505 Energy switch 1 GND 2 GND 3 ENERGYSW 4 ENOUTB 5 ENOUTA 6 5V SW506 Adjust Set switch 1 GND 2 GND 3 SETSW 4 ADOUTB 5 ADOUTA 6 5V SW507 Power switch 1 LINE 2 NEUTRAL 3 SWLINE 4 SWNEUTRAL 6 REAR PANEL CN601 9pin DB 1 RS232 2 RDX 3 TXD 4 DTR 5 GND 6 7 RTS 8 CTS 9 CN602 3pin IEC 1 line Power 2 neutral 3 gnd ...

Страница 31: ...SOMATICS LLC REV 3 1 09 15 2015 31 SW602 115 230 1 TR1 Voltage switch 2 TR2 3 TR3 4 SW601 Alarm test 1 ALARMSW 2 GND ...

Страница 32: ...0 100 50 70 70 50 100 50 75 70 60 110 60 80 70 60 120 60 85 70 60 120 60 90 70 70 130 70 95 70 70 140 70 100 70 70 140 70 Pulse width ms for Thymatron System IV US version PercentDGx Lowest Low 25 Low 5 Energy Charge Charge Charge Charge 5 1 0 0 25 0 25 0 5 10 1 0 0 25 0 25 0 5 15 1 0 0 25 0 25 0 5 20 1 0 0 25 0 25 0 5 25 1 0 0 25 0 25 0 5 30 1 0 0 25 0 25 0 5 35 1 0 0 25 0 25 0 5 40 1 0 0 25 0 25...

Страница 33: ...ailable in the US PercentPulsewidth Frequency Energy ms Hz 5 0 5 10 10 0 5 10 15 0 5 20 20 0 5 20 25 0 5 20 30 0 5 30 35 0 5 30 40 0 5 30 45 0 5 40 50 0 5 40 55 0 5 40 60 0 5 50 65 0 5 50 70 0 5 50 75 0 5 60 80 0 5 60 85 0 5 60 90 0 5 70 95 0 5 70 100 0 5 70 110 0 75 60 120 0 75 60 130 0 75 70 140 0 75 70 150 0 75 70 160 1 0 60 170 1 0 60 180 1 0 70 190 1 0 70 200 1 0 70 ...

Отзывы: