background image

SRU User Guide

Page 7

Idle

 

Status:

 indicates whether the SRU is idle, logging, or downloading/viewing data logs.

 Battery icon

: represents the battery life remaining for the SRU. Battery life is also numerically displayed 

as a percentage to the left of the icon. 

Note:

 It is recommended to replace the battery when the level falls below 20%.

Arrows

: are used to navigate through and select the SRU menu items. Pressing and holding any of the arrow 

buttons for two seconds, turns the SRU on/off.

Figure 2-2  SRU LCD Display

2.2  LCD Display and Navigation

2.0  SRU Operation

2.1  Start Up

Press and hold any arrow button for two seconds to turn the SRU on.  When starting, the SRU will perform 

a self-test. The start-up screen is shown first; following a short pause the main menu will be displayed. The 

SRU has an auto-off after 2 minutes of inactivity.  

Figure 2-1  Start-up Screen

Use the up/down arrows to scroll through menu items.

Use the left/right arrows to select the menu actions shown at the bottom left/right 

of the LCD (e.g. Next, Back, etc.)

Next

Back

Level (Comp):       10.045m

Temperature: 28.234ºC

Conductivity:     125.4µS

Idle

99%

Next

Take a Reading

Download/View Files

Status/Setup

Idle

99%

Next

On

Back

Turn Compensation:

Baro Reading:    98.89kPa

Idle

99%

Back

Start

12 mins

Sample Rate:

Compensation:    On

Idle

99%

Stop

Save

Level (Comp):     10.045m

Temperature: 28.234

ºC

Conductivity:     125.4

µ

S

Log 56

99%

Home

Data Saved

Compensated

Barometrically

Log 57

99%

Home

Data Saved

Uncompensated

Log 57

99%

Stop

Save

Level (Uncomp):    10.045m

Temperature: 28.234

ºC

Conductivity:     125.4

µ

S

Log 56

99%

Back

Start

12 mins

Sample Rate:

Compensation:    Off

Idle

99%

Off

Next

Take a Reading

Download/View Files

Status/Setup

55%

Downloading

Download

99%

99%

Cancel

Downloading Logs:

Next

Take a Reading

Download/View Files

Status/Setup

Idle

99%

Next

Back

LCD Contrast

SRU Status

SRU Status

Logger Status

Idle

99%

Next

Back

LCD Contrast

Logger Status

SRU Status

Idle

99%

Download Logger

Idle

99%

Idle

99%

View Files

Next

Back

Download Logger

View Files

Idle

Next

Back

99%

Next

Back

LCD Contrast

Logger Status

Idle

99%

Home

Logger Data

Saved to file

1081225-01.10.21-12.30.xle

Next 3

Home

3/11 Files

99%

1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U

Next

Back

Idle

99%

Logger Serial:     1081234

Logger Memory:     50%

Logger Battery:    85%

Next

Back

Idle

99%

Log Mode:      Schedule

Logging Rate:     15 secs

Logger Status:     Started

Back

Home

Idle

99%

Location:     Wellhead 234

Name:     Zanzibar 1

Firmware:     V1.004

Back

Idle

99%

SRU Serial:     123456

SRU Memory:     50%

Firmware:     V1.000

Back

Home

Idle

99%

55%

Contrast:

Set LCD

Содержание SRU 4001

Страница 1: ...Model 4001 Solinst Readout Unit SRU User Guide September 15 2021...

Страница 2: ......

Страница 3: ...10 4 0 Real Time Readings 11 4 1 Take a Reading 11 4 2 Barometric Compensation 11 4 3 Real Time Logging 12 5 0 Downloading and Viewing Data Files 13 5 1 Download Logger 13 5 2 View Data Files 13 6 0 D...

Страница 4: ......

Страница 5: ...ogger Software The SRU provides access to useful information about the connected datalogger s status including battery level memory serial number firmware version and logging mode You can also access...

Страница 6: ...004 AquaVent 5 1 004 Levelogger Edge 3 004 Barologger Edge 3 004 Levelogger Junior Edge 3 004 LTC Levelogger Edge 1 003 Rainlogger Edge 3 001 LevelVent 1 000 AquaVent 1 000 It is also important to en...

Страница 7: ...Figure 2 2 SRU LCD Display 2 2 LCD Display and Navigation 2 0 SRU Operation 2 1 Start Up Press and hold any arrow button for two seconds to turn the SRU on When starting the SRU will perform a self te...

Страница 8: ...xt Back Download Logger View Files Idle Next Back 99 Home Logger Data Saved to file 1081225 01 10 21 12 30 xle Next 3 Home 3 11 Files 99 1081225 01 10 21 12 30 D 1081225 01 15 21 15 56 C 1081225 01 16...

Страница 9: ...Download View Files Status Setup Idle 99 Next Back LCD Contrast SRU Status SRU Status Logger Status Idle 99 Next Back LCD Contrast Logger Status SRU Status Idle 99 Download Logger View Files Next Bac...

Страница 10: ...L5 Threaded or Slip Fit Adaptor can also be used to connect a Levelogger to the SRU cable 3 2 LevelVent Connection The SRU cable connects to a LevelVent s Wellhead Hold the Wellhead and thread the SR...

Страница 11: ...ressure sensors therefore barometric effects are already eliminated from the level reading through the vented cable Barometric compensation can be turned off in the same screen Next Back Level Comp 10...

Страница 12: ...99 Stop Save Level Comp 10 045m Temperature 28 234 C Conductivity 125 4 S Log 56 99 Home Data Saved Compensated Barometrically Log 57 99 Home Data Saved Uncompensated Log 57 99 Stop Save Level Uncomp...

Страница 13: ...load Logger View Files Idle Next Back 99 Home Logger Data Saved to file 1081225 01 10 21 12 30 xle Next 3 Home 3 11 Files 99 1081225 01 10 21 12 30 D 1081225 01 15 21 15 56 C 1081225 01 16 21 22 50 U...

Страница 14: ...1081234 Logger Memory 50 Logger Battery 85 Back I Log Mode Logging Rate Logger Status Back Idle 99 SRU Serial 123456 SRU Memory 50 Firmware V1 000 Back Home Idle 99 55 Contrast Set LCD Next s Idle 99...

Страница 15: ...00 x 600 Ports USB or RS 232 Web Download 1 Download the newest version of Levelogger Software by visiting https www solinst com downloads 2 If the installer does not automatically start to activate t...

Страница 16: ...RU to PC Connection To start the Levelogger Software click or click the Start button and select Programs Solinst Levelogger 4 Levelogger 4 6 1 Go to the SRU tab and select the appropriate Com Port for...

Страница 17: ...SRU User Guide Page 17 Select the file s from the list that you want to download and click the Download Data icon Figure 7 4 Levelogger Software SRU Tab...

Страница 18: ...r Guide Page 18 7 3 View Data Data files can be viewed in the Data Control tab by selecting Finish and Open all or select Finish if you do not need to view the data right away Figure 7 5 Finish Data D...

Страница 19: ...SRU tab Note An error message on the SRU screen will indicate when its memory is full You can also view the status of the SRU memory using the Statu Status menu see Section 2 5 7 5 Synchronize SRU Tim...

Страница 20: ...SRU Battery Installation 8 2 Software and Firmware Updates Each time the software is opened it automatically checks on line for software updates If there is an update Software Update Available will ap...

Страница 21: ...the Com Port that the SRU is connected from the drop down menu 2 Click the Open button which should open a file dialog asking for the firmware file ssf to upload Navigate to the directory where the fi...

Страница 22: ...5 Downloading Download 99 99 Cancel Downloading Logs Next Take a Reading Download View Files Status Setup Idle 99 Next Back LCD Contrast SRU Status SRU Status Logger Status Idle 99 Next Back LCD Contr...

Страница 23: ...Canada Ltd 35 Todd Road Georgetown ON L7G 4R8 Tel 1 905 873 2255 800 661 2023 Fax 1 905 873 1992 instruments solinst com www solinst com High Quality Groundwater and Surface Water Monitoring Instrume...

Отзывы: