background image

Precision Thermal Control for Optimal Laser 
Performance and Stability

THERMORACK 800

Product Manual

Содержание THERMORACK 800

Страница 1: ...Precision Thermal Control for Optimal Laser Performance and Stability THERMORACK800 Product Manual...

Страница 2: ...2 OPERATION _______________________________________________________________________ 12 5 1 Simple Operation _____________________________________________________________ 12 5 2 Advanced Operation ____...

Страница 3: ...ur sole responsibility that the ThermoRack 800 All Models meets the provisions of the directives 2004 108 EC EMC Directive 2006 95 EC Low Voltage Directive EN 61326 1 2006 Emissions and Immunity EN 61...

Страница 4: ...tor only trained personnel may interact with internal components Caution Heavy object System weighs 55 lbs 25 kg Two people are required to lift CAUTION Never disassemble the chiller as irreparable da...

Страница 5: ...The red WARNING equilateral triangle symbol appears throughout the manual accompanying certain maintenance and repair activities Please follow the important instructions accompanying this symbol to a...

Страница 6: ...3 5 liters per minute of constant temperature coolant with PID control for both cooling and heating With fewer moving parts the system is highly reliable and energy efficient The ThermoRack 800 has b...

Страница 7: ...ms with centrifugal pumps as water absorbs CO2 from the air causing its pH to become acidic Acidic water dissolves copper which then plates out inside the magnetic drive of the centrifugal pump causin...

Страница 8: ...ID STATE COOLING SYSTEMS 167 MYERS CORNERS ROAD WAPPINGERS FALLS NY 12590 7 TELEPHONE 845 296 1300 FAX 845 296 1303 WEB WWW SSCOOLING COM VERSION M8 Figure 1 ThermoRack 800 Cooling Capacity in 20 C Am...

Страница 9: ...6 1300 FAX 845 296 1303 WEB WWW SSCOOLING COM VERSION M8 SECTION 3 HOOK UP _______________________________________________________ Figure 3A Front View Figure 3B Rear View with 5X20 10A fast acting fu...

Страница 10: ...ance is required in front and in back to ensure adequate airflow 3 2 ELECTRICAL CONNECTIONS SEE FIGURES 3A AND 3B Power The ThermoRack 800 AC power inlet is an IEC320 C20 socket Plug the line cord pro...

Страница 11: ...k adding coolant as required to maintain the tank level just below maximum 4 Replace tank cap Recommended Coolants Solid State Cooling Systems recommends using Koolance a pre mixed 27 propylene glycol...

Страница 12: ...hiller on via the switch located on the front Fluid should begin flowing into the bucket If it does not turn off the chiller wait 30 seconds then turn the chiller back on Once fluid begins flowing tur...

Страница 13: ...operating turns off temperature control 5 1 SIMPLE OPERATION The ThermoRack 800 comes with preset operating parameters that will work well for most applications If temperature control at one temperatu...

Страница 14: ...several user adjustable parameters To access this menu press and hold the ENTER button for 3 seconds and then enter password 0000 using the UP DOWN and ENTER keys This menu allows access to the follo...

Страница 15: ...on scale up 2 C to match the external sensor TEMPUNIT Sets temperature units in degrees Celsius or Fahrenheit BKLIGHT Setting this parameter to ON turns on the display back light setting this paramete...

Страница 16: ...t the front and rear air inlet and outlet gratings are not blocked If airflow is not blocked contact SSCS for an RMA number to return the unit for fan replacement Pump Fail The liquid heat exchanger p...

Страница 17: ...Format ASCII Insert a carriage return 0x0Dhex at the end of each command string A carriage return 0x0Dhex is sent at the end of each response Table 2 USB Commands PARAMETER GET Data Command Response...

Страница 18: ...3 Cr 69 2 Cr 119 Cr 7 Cr 0 Cr ETX STAT1A Bit Field Definition convert decimal number sent into binary read as follows __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ ___Chiller Running 1 RUNNING 0 STO...

Страница 19: ...inal Equipment DTE Table 1A Signal definition and wiring for Host wired as DTE Computer Host Master Male 9 Pin D sub Pin Chiller Slave Female 9 Pin D sub Pin 2 Receive RXD 2 Transmit TXD 3 Transmit TX...

Страница 20: ...erature PUMPTEMP XX X Actual TE PWM PWM XX X Status word 1 5 ASCII bytes STAT1A 0 65535 Faults word 1 5 ASCII bytes FLTS1A 0 65535 Reset All Alarms and Restart RESTART GETSET2 parameters GETSET2 retur...

Страница 21: ...NG 0 COOLING ___________________ genwarning 1 WARNING 0 NONE ______________________ genalarm 1 ALARM 0 NONE _______ rsvd ____ __________________________ autotune 1 ON 0 OFF FLTS1A Bit Field Definition...

Страница 22: ...xterior surfaces of the ThermoRack 800 may be cleaned with a non shedding wipe dipped in isopropyl alcohol SECTION 9 TECHNICAL SUPPORT _____________________________________________ Delighting our cust...

Страница 23: ...ized System of Classification and Labeling of Chemicals GHS Physical Hazard Not applicable Health Hazard Skin Irritation Category 2 Eye Irritation Category 2 Environmental Hazard Not applicable Label...

Страница 24: ...on and immediately contact emergency medical assistance If ingested Never give anything by mouth to an unconscious person Rinse mount with water and consult a medical doctor immediately Other medical...

Страница 25: ...age KOSHA No data available US ACGIH No data available Appropriate engineering controls Respiratory protection Approved respirator equipped with cartridge for organic vapors Eye protection Protective...

Страница 26: ...ratory sensitization No data available Skin sensitization No data available Germ cell mutagenicity No data available Carcinogenicity Not classifiable from IARC EC ESIS Reproductive Toxicity No data av...

Страница 27: ...S US NLM Toxic and ecological information OECD SIDS IUCLID US NLM IARC EC ESIS CCRIS Acronyms and Websites EC ESIS European chemical Substances Information System http esis jrc ec europa eu IUCLID Int...

Страница 28: ...use Details of the supplier of the safety data sheet Company Fisher Scientific One Reagent Lane Fair Lawn NJ 07410 Tel 201 796 7100 Emergency Telephone Number CHEMTREC Inside the USA 800 424 9300 CHEM...

Страница 29: ...sposal Dispose of contents container to an approved waste disposal plant Hazards not otherwise classified HNOC WARNING Reproductive Harm https www p65warnings ca gov 3 Composition Information on Ingre...

Страница 30: ...breathing apparatus pressure demand MSHA NIOSH approved or equivalent and full protective gear NFPA Health Flammability Instability Physical hazards 2 1 1 N A 6 Accidental release measures Personal P...

Страница 31: ...nd safety practice 9 Physical and chemical properties Physical State Viscous liquid Liquid Appearance Colorless Odor Odorless Odor Threshold No information available pH 5 5 7 5 50 aq sol Melting Point...

Страница 32: ...ot listed Not listed Not listed Mutagenic Effects No information available Reproductive Effects No information available Developmental Effects No information available Teratogenicity No information av...

Страница 33: ...Section 5 e Consent order under TSCA F Indicates a substance that is the subject of a Section 5 f Rule under TSCA N Indicates a polymeric substance containing no free radical initiator in its invento...

Страница 34: ...e Pollutant N U S Department of Homeland Security This product does not contain any DHS chemicals Other International Regulations Mexico Grade Slight risk Grade 1 16 Other information Prepared By Regu...

Страница 35: ...ent installation storage or usage occurring during a situation specifically cautioned against or noted in the product manual Specific situations which invalidate the warranty include but are not limit...