EN
30
INTRODUCTION
Metrological maintenance and care information
•
Do not perform any metrological maintenance
if the device is in use.
•
For reliable operation, we recommend that
the device is subject to metrological control
once every two years pursuant to Sec. 14 of
the Medical Product Operating Directive (MP-
BetriebV). We recommend testing the device’s
performance once every two years and after
metrological control and repair, at a minimum
by testing the tolerance requirements of the
pressure display and tightness of the cuff
(testing at least at 50 mmHg and 200 mmHg).
•
The metrological control must be performed by
a licensed tester or testing institute.
•
We recommend cleaning the device and cuff
only with a soft cloth.
Storage information
•
If the device is not being used, store it in a dry
location away from heat, dust, fibres, and out of
direct sunlight. Do not place heavy objects on
the storage box.
Information about handling batteries
•
Do not mix new and used batteries.
•
Do not use different types of batteries at the
same time.
•
Do not throw batteries into a fire. Batteries may
explode or leak. Remove the batteries if the
device is not used for a long time.