Acid Cleaning Procedure:
Before cleaning, measure and record the flow and pressure drop through the
electrode and concentrate streams.
Close (seal) the Feed and Product ports to keep acid from contaminating the ion
exchange resins.
Connect the discharge side of the acid-capable pump to (
elec in
and
conc in
)
inlet ports. Do not feed acid to the dilute/purifying chambers or the resins will
require a very long regeneration time. If the acid is fed to the product chambers,
rinse time will be 6-8 hours, and it may take 16 hours to reach equilibrium quality.
Connect the suction side of the pump to a 20-50 liter (5-10 gallon) plastic tank
filled with one of the above acid cleaning solutions.
Direct (
elec out
and
conc out
) outlet ports to a waste recovery tank suitable for
acid solutions, or to the chemical feed tank.
Pump the acid slowly through the EDI module until acid and waste products
come out the outlet, then turn off the pump and allow the cleaner to soak in the
module for 5 minutes or longer. Dense scale deposits may require longer soaks
or stronger acid.
Reactivate the pump periodically until the scale is dissolved.
Avoid pumping air
into module.
Fill the tank with deionized (DI) water and pump it through the module to rinse out
the cleaner, with the outlets going to drain. Continue to add rinse water until the
effluent measures no greater than 5 ppm in TDS, and the pH is within normal
operating range.
Disconnect the lines from the ports, and connect the ports to their original
plumbing. Measure the pressure drop, flow, pH, and TDS, and record. The
pressure drop should have decreased into the normal range.
Run the EDI in regeneration mode until an ion material balance is achieved. Run
the EDI in standard mode until water quality is re-established. If the purifying
chambers are contaminated with acid, it will require a rinseout time of at least 6
hours, and ultimate product quality will be reached after about 16 hours of
operation.
Содержание Electropure EXL Series
Страница 51: ...SNOWPURE LLC 2005 2018 VERSION 3 5 XL EXL FEBRUARY 2018 PAGE 51 Figure 5 EXL EDI Module Drawing Dimensions...
Страница 53: ...SNOWPURE LLC 2005 2018 VERSION 3 5 XL EXL FEBRUARY 2018 PAGE 53...
Страница 54: ...SNOWPURE LLC 2005 2018 VERSION 3 5 XL EXL FEBRUARY 2018 PAGE 54 Figure 8 Manifolding Options...
Страница 60: ...SNOWPURE LLC 2005 2018 VERSION 3 5 XL EXL FEBRUARY 2018 PAGE 60 EC Declaration of Conformity...
Страница 71: ...SNOWPURE LLC 2005 2018 VERSION 3 5 XL EXL FEBRUARY 2018 PAGE 71 Appendix 7 DataFormforElectropure XL EDIModule...