1
1. Expected use
This product is used in various surgical lighting in hospital operating room. It can
also be use in medical lighting situations in emergency room.
2. Conditions of use
2.1 Normal working conditions
Temperature: 10℃~30℃;
Relative humidity: ≤70%;
Atmospheric pressure range: 860hPa~1060hPa;
This equipment can not be worked in flammable and explosive gas environment.
The type of device security: I type equipment;
Working system: continuous operation;
Liquid injection protection level: IPXO.
2.2 Power parameters
Power supply voltage: AC220(±10%)V;
Frequency: 50Hz;
Input power: 70VA;
Main lamp power: 70W.
2.3 suitable mounting height: 3200~3300 mm
Low-storey may bring inconvenience to users. please refer to the product diagram.
in special circumstances, you should contact with the manufacturer when purchasing.
3.Technical specifications
Type
JM-LED01 Main lamp
Diameter of Light Head (mm)
500
Quantity of LEDS
48
Illumination (10 Steps Adjustable)
40~160klx
Color rendering index
Ra>95
Color temperature (5 Steps Adjustable)
3500-5000K
Depth of illumination (mm)
1000
Spot diameter
120-250mm
Endoscopy Mode
Available
Endoscopy Illumination
40,000Lux
LED Life
>50,000 Hours
The model of LED
OSRAM GWCS8PM1.CM
UPS
Available
Control Panel
LCD
www.smitsonjupiter.com