Safety alarms
General
All alarms for RENASYS™ GO are determined to be "Low Priority," indicating that operator awareness is required
(IEC 60601-1:2005, 3rd edition and IEC 60601-1-8:2006).
In an alarm state with the audio paused, if a different alarm state occurs, the audio pause will be canceled, the
audio will sound and the display screen will indicate the new alarm state. When multiple alarm states are present,
the device's display screen will alternate between the display information for each alarm activated.
Cautions – Lack of alarms:
RENASYS GO device is equipped with high flow/leak, low vacuum and complete blockage/canister over-capacity
alarms to alert users to these events occurring during therapy. These alarms are designed to activate based on
changes in pressure status detected by device. However, there are scenarios that may occur during therapy that
can impact alarm functionality. Therefore, it is important that patient, device and wound dressing are monitored at an
appropriate interval to ensure therapy is being delivered.
Over vacuum alarm:
The system has encountered an excessively high vacuum (of >255mmHg) and device will stop delivering therapy.
Event must be investigated by healthcare provider immediately and corrected.
Alarm state
Visual indication
Display screen
Audible alarm
Over vacuum
Status/alarm indicator:
illuminates solid yellow
! THERAPY STOP
! OVER VACUUM
Beeps 2 times every 20
seconds
Audio cannot be paused
High flow/leak alarm:
Device detects a significant leak in the system. Once corrected, alarm will automatically reset and
status/alarm indicator will return to solid green.
Alarm state
Visual indication
Display screen
Audible alarm
High flow/leak
Status/alarm indicator:
illuminates solid yellow
! WARNING
! LEAK
Beeps 2 times every 20
seconds
Audio may be paused
High flow/leak
Audio paused
Status/alarm indicator:
illuminates solid yellow
! AUDIO PAUSED
! LEAK
Paused for 2–3 minutes
Low vacuum alarm:
Vacuum level is lower than set point of NPWT by >15mmHg. Once corrected, alarm will automatically reset and
status/alarm indicator will return to solid green.
Alarm state
Visual indication
Display screen
Audible alarm
Low vacuum
Status/alarm indicator:
illuminates solid yellow
! WARNING
! LOW VACUUM
Beeps 2 times every 20
seconds
Audio may be paused
Low vacuum
Audio paused
Status/alarm indicator:
illuminates solid yellow
! AUDIO PAUSED
! LOW VACUUM
Paused for 2–3 minutes