To verify the recirculating solution temperature, you will need the
following:
• TEMP CHECK (HLTA-390)
• HOTLINE
3 Warmer
To Verify the Recirculating Solution Temperature:
1
Plug the
HOTLINE
3
Warmer into a power outlet.
2
Place the TEMP CHECK on the top right corner of the
HOTLINE
3
Warmer and plug it into the socket on the right side
of the
HOTLINE
3
Warmer.
3
Remove the black label from the auxiliary outlet on the back of the
HOTLINE
3
Warmer and plug in the TEMP CHECK power cord.
Note: The auxiliary outlet is for use only with Smiths Medical
accessories.
4
Turn ON the
HOTLINE
3
Warmer. Allow 15 minutes for the
temperature to stabilize.
5
If the TEMP CHECK display indicates a temperature between
37
°
C and 38
°
C, and the
HOTLINE
3
Warmer display equals the
TEMP CHECK display, recirculating solution verification is
complete. Refer to the TEMP CHECK Manual for OVERTEMP
ALARM verification.
6
If the TEMP CHECK display does not indicate a temperature
between 37
°
C and 38
°
C, refer to the TEMP CHECK Manual for
calibration instructions.