S E C T I O N 7
•
O p e r a t i n g I n s t r u c t i o n s
HOTLINE
®
3 Blood and Fluid Warmer | Operator’s Manual
19
b
Push on the center (
c
) of the connector until it clicks in the
socket (
d
).
3
Turn ON the power switch.
• The green Operating LED on the Display Panel illuminates.
• The recirculating solution temperature display will begin to
increase.
• The recirculating solution path in the HOTLINE
3 Fluid
Warming Set will automatically prime.
4
Remove the end cap and inspect the patient end of the
HOTLINE
3
Fluid Warming Set for leaks to confirm the integrity
of the intravenous pathway.
Step 3 - Connect the Intravenous Administration Set
WARNINGS
WARNINGS
• Remove all air from the HOTLINE
®
3 Fluid Warming Set, L-10,
PC-8, and YC-8 before connecting to the patient. Failure to do
so may result in introduction of air to the patient, which may
contribute to serious patient injury or death.
• Do not stick the HOTLINE
®
3 Fluid Warming Set with needles, as
this will breach the I.V. path and compromise the integrity of the
patient intravenous line. If a Disposable Set with a breached
recirculating solution path/intravenous path is used, then
patient illness may occur because of the HOTLINE
®
3 Warmer's
recirculating solution entering the patient's blood stream.
1
Connect the I.V. fluid and the intravenous administration set to the
HOTLINE
3
Fluid Warming Set.
2
Fully prime the intravenous administration set, the
HOTLINE
3
Fluid Warming Set, and patient extension set (if used).
o
o
C
C
d
o
o
C
C
c