42
EN
8. Cleaning, maintenance and expected life
Cleaning
As you use the RENASYS™ GO device, your healthcare provider should
clean it on a regular basis to ensure good hygiene. The device may be
wiped clean with a diluted bleach solution (5mL bleach: 1L water), wiped
again with a damp cloth using clean water, and dried with a clean cloth to
ensure the device does not remain wet.
Maintenance and expected life
Your healthcare provider will ensure the device has been properly
maintained prior to use. Your device should last the duration of your
therapy. If a replacement is needed, your healthcare provider will provide
the items needed.
9. Technical specifications
Environmental conditions
Storage and Transport
14 to 131°F / -10 to 55°C
Operational Temperature
41 to 95°F / 5 to 35°C
Relative Humidity
30% to 70% RH
Atmospheric Pressure
700 mbar to 1,060 mbar
Ingress protection: IP 22
The device housing is designed to protect its internal components from
a moderate level of external debris or small objects and falling water
(e.g. light rain).