27
EN
Alarm/Indicator:
Device status:
Indicator light/
audio:
Impact to therapy:
Steps to remedy:
Device is not plugged into an
electrical outlet and battery
is low.
Status/Indicator:
Solid Yellow
The audible alarm
will sound
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
can be paused
by pressing the
Audio Pause
button. When
pressed the light
above the button
will illuminate.
Device will continue
to operate normally
for
3 hours
and will
stop operating once
the battery is fully
discharged.
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.
Device is not plugged into an
electrical outlet and battery
is very low.
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
will sound.
The audible
alarm cannot be
paused.
Device will continue
to operate normally
for
1 hour
and will
stop operating once
the battery is fully
discharged.
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.