background image

13

10600266 Rev. E

DYONICS™ Handpieces     

Instructions for Use

Service

Cleaning Verification

1.  After cleaning, inspect devices under normal lighting 

to ensure that all visible soil has been removed.

2.  If not visibly clean, repeat cleaning and reinspect.
3.  For difficult-to-view areas, 3% hydrogen peroxide 

may be applied (bubbling is evidence of the 

presence of blood).

 

Note:

 Rinse instruments thoroughly with warm tap 

water following any hydrogen peroxide testing.

Storage

Medical devices that will be stored between cleaning 

and sterilization should be dried with a low-linting,  

non-abrasive soft cloth to prevent microbial 

contamination that could result from wet instruments. 

Devices should ALWAYS be thoroughly cleaned prior  

to storage. 

Sterilization Parameters

DYONICS™ Handpieces must be sterilized by steam, 

using one of the following methods only:
 

High Temperature Prevac Steam:

 Steam, 

pre-vacuum, at 270°–275° F (132°–135° C)  

for 4 minutes.

 

High Temperature Gravity Steam:

 Steam, 

gravity method, at 270°–275° F (132°–135° C)  

for 10 minutes.

 

World Health Organization (WHO) Steam Cycle:

 

Steam, 273°–280° F (134°–138° C) for 18 minutes.

CAUTION: Do not cool the handpiece by immersing it in 

cold water.
CAUTION: To prevent damage to the control unit 

connector ports, do not plug in wet. Ensure that cleaned 

or sterilized cable connectors are completely dry prior 

to connecting to the control unit.

Service

Repairs and adjustments are to be performed only  

by Smith & Nephew authorized service centers. 
If service becomes necessary, call Smith & Nephew 

Customer Service prior to returning the device and 

request a Return Authorization (RA) number. Customer 

Service can also explain the available Service 

Replacement and Repair Programs. 
Service items should be carefully disinfected and 

packaged and returned postpaid to Smith & Nephew. 

Smith & Nephew Customer Service can provide 

additional instructions. 
Product returned that has been serviced by an 

unauthorized third party repair shop and/or sterilized 

with a sterilization method other than one approved 

by Smith & Nephew and/or damaged by contact with 

other mechanical devices (such as shavers) may incur 

additional costs, regardless of warranty status. 
It is not necessary to include accessory items (i.e., 

power cords, etc.) when returning a device for service.
There are no user-serviceable components, 

adjustments, or repairs other than those specifically 

described in this manual. 

Handpiece Cable Replacements

The following are replacement parts for the  

DYONICS™ Handpieces: 

REF

Description

7208207

Shaver handpiece cable 

7209071

Ultralight Motor Drive cable 

7208208

Mini Motor Drive cable 

The handpiece cable replacement is supplied in a 

kit with a set of installation instructions. A standard 

adjustable wrench and alcohol wipes will be required to 

clean the connector pins. 

Note: 

The Smith & Nephew DYONICS POWERMAX™, 

POWERMAX ELITE Handpieces (REF 7209891, 

REF 72200616, REF 72200617, and REF 7210542) and 

DYONICS POWERMINI™ (REF 7221500 and REF 7221503) 

do not have field-replaceable cables, and must be  

returned to Smith & Nephew for cable replacement. 

To replace the handpiece cable: 

1.  Using a standard adjustable wrench, loosen the 

stainless steel locking nut at the proximal end of  

the handpiece. Remove the nut from the old cable. 

2.  Ensure that the REF number on the replacement 

cable is identical to the REF number on the old cable.

3.  Clean and dry the connector pins and baseplate  

of the handpiece. 

4.  Carefully align the pins to the sockets on the 

replacement cable and join the two pieces. 

5.  Tighten the locking nut securely to ensure a good 

electrical ground connection. 

6.  Connect the handpiece to the control unit and  

verify operation of the handpiece using the new 

power cable. 

Note:

 The Mini Motor Drive cable and the Ultralight 

Motor Drive cable are very similar. The threads on the 

locking nut are different for each type of cable. If cable 

replacement is difficult, recheck the REF number to 

ensure it is the correct cable.

Содержание 7205354

Страница 1: ...DYONICS Handpieces Instructions for Use 10600266 Rev E ...

Страница 2: ......

Страница 3: ...r on the order of a physician EU Not for general waste Humidity Forward clockwise direction of rotation distal view Reverse counterclockwise direction of rotation distal view Oscillate alternating forward reverse direction of rotation Suction On fluid flow through handpiece unrestricted Suction Off fluid flow through handpiece blocked ...

Страница 4: ...indications 5 Warnings 6 Precautions 6 Instructions for Use 8 Suction 8 Hand Control Handpieces 9 Non Hand Control Handpieces 9 Installing a Blade in a Handpiece 11 Installing a Blade in the Mini Motor Drive 11 Blade Window Lock 11 Forward Reverse Speed Shift 11 Cleaning and Sterilizing 12 Cleaning 12 Cleaning Verification 12 Storage 12 Sterilization Parameters 12 Service 13 Handpiece Cable Replac...

Страница 5: ...blades for resection of soft and osseous tissue including but not limited to use in large articular cavities small articular cavities and for Functional Endoscopic Sinus Surgery FESS The FESS application is limited to small blades that are appropriate for the procedure Contraindications Smith Nephew DYONICS Handpieces and Blades should not be used with patients exhibiting ankylosis and also lackin...

Страница 6: ...avoid patient burns do not place the handpiece on the patient when it is not in use Precautions U S Federal law restricts these devices to sale by or on the order of a physician Prior to each use inspect the device to ensure it is functioning properly and not damaged Do not use a damaged device In all cases the judgment of the physician in the best interest of the patient may take precedence over ...

Страница 7: ...cted by one or more of the following measures Reorient or relocate this equipment the other equipment or both Increase the separation between the pieces of equipment Connect the pieces of equipment into different outlets or circuits Consult a biomedical engineer Environmental Protection This equipment contains electronic printed circuit assemblies At the end of the useful life of the equipment it ...

Страница 8: ... for replacement The DYONICS POWERMINI Handpiece is not compatible with the DYONICS POWER Shaver System Suction is connected to a portal on the proximal end of the handpiece and removes fluid and debris drawn through the blade window In addition to removing fluids the suction cools and lubricates the motor drive Cutting takes place when the blade edge rotates across the blade s outer window The bl...

Страница 9: ... 7205357 REF 7205971 and REF 72201503 are hand held motor drives that are electrically connected to the control unit via a 10 foot 3 meter cable The non hand control handpieces are activated by a footswitch These handpieces run a variety of DYONICS Disposable Endoscopic Blades The handpieces are designed to sense the blade type installed and communicate that information to the control unit which s...

Страница 10: ...PM Range Default Range Default Range Default Range Default DYONICS POWER REF 7205354 7205355 Long Blades Straight and Curved TURBOWHISKER 100 3000 2000 100 3000 2000 500 3000 2000 1 9 1 Rev 1 8 2 Rev 8 2 Rev DYONICS POWERMAX REF 7210542 7210586 7209891 3 5 mm Curved Blade 100 3000 2000 100 3000 2000 500 3000 2000 1 9 1 Rev 1 8 2 Rev 8 2 Rev DYONICS POWERMAX ELITE REF 72200616 72200617 Curved HELIC...

Страница 11: ...sition of the inner blade relative to the opening at the top of the outer blade Hand Control Operation Press and hold the Oscillate button on the hand control handpieces for one second Footswitch Operation Press and hold the Window Lock button on the footswitch to rotate the blade When the blade is in the best stopping position for the procedure release the Window Lock button Control Unit Operatio...

Страница 12: ...ush in the device spigot proximal end of the device and clean the lumen using twisting and in out movements of the brush for a minimum of 30 seconds Reinsert the lumen brush from the distal end of the device end and clean the lumen from the opposite direction using twisting and in out movements of the brush for a minimum of 30 seconds Using the blade port brush clean the nose cavity using twisting...

Страница 13: ...uct returned that has been serviced by an unauthorized third party repair shop and or sterilized with a sterilization method other than one approved by Smith Nephew and or damaged by contact with other mechanical devices such as shavers may incur additional costs regardless of warranty status It is not necessary to include accessory items i e power cords etc when returning a device for service The...

Страница 14: ...s Forward Reverse Oscillate Blade Window Lock and Speed Shift Equipped with 10 foot 3 meter autoclavable power cord not field replaceable Suction control lever DYONICS POWERMAX and POWERMAX ELITE Handpieces REF 7210542 REF 72200616 REF 7220617 Length 7 1 18 0 cm Weight 7 5 oz 223 grams Lightweight handpiece with push button motor controls Forward Reverse Oscillate Blade Window Lock and Speed Shift...

Страница 15: ...eight 5 0 oz 142 grams Standard handpiece with push button motor controls Forward Reverse Oscillate Blade Window Lock Equipped with 10 foot 3 meter autoclavable power cord not field replaceable Suction control lever Allows for shaver blades to be inserted with the blade window aligned to the suction control lever or 180 opposite the lever Specifications subject to change without notice Handpiece d...

Страница 16: ... For a Return Authorization RA number or for additional information on this program call Customer Service at 1 800 343 5717 in the U S or contact your authorized representative 24 hour shipment is not offered in all countries Repair Service Program For devices no longer under warranty repairs can be made by Smith Nephew or by an authorized agent Non warranty repairs will be made at the list price ...

Отзывы: