Z/ISE3-TFK26GB-A
5. Functions Setting
1. Set value input mode
Press the "SET" button to display "P1-20 (*1)".
The output OUT1 (1) set value input mode is selected.
(*1: If the set value of P1 is -20).
2. OUT1 (1) set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
3.
Press the "SET" button to save the set value and select OUT1 (2)
set value input mode. The set value of P2 is displayed.
4. OUT1 (2) set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
5.
Press the "SET" button to save the set value and select OUT2 (1)
set value input mode.
The set value of P3 is displayed.
6. Output OUT2 (1) set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
7.
Press the "SET" button to save the set value and select OUT2 (2)
set value input mode.
The set value of P4 is displayed.
8. Output OUT2 (2) set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
9.
Press the “SET” button to save the set value and exit this mode.
2-output Type
Note)
P1: Set value for OUT1 (1) P2: Set value for OUT1 (2)
P3: Set value for OUT2 (1) P4: Set value for OUT2 (2)
Note)
• Hysteresis mode (same as for positive pressure use)
When the value of hysteresis is set to 2 digits or less, the switching
output might chatter due to fluctuation of the input pressure around
its set point.
•
Window comparator mode (same as for positive pressure use)
Since the hysteresis will be 3 digits, separate P1 from P2 (in case of
2-output type, same as for P3 and P4) by 7 digits or more.
* 1 digit is the minimum setting unit.
Output Method
Hysteresis mode (P1 P2, P3 P4)
Window comparator mode (P1<P2,P3<P4)
P1
(P3)
P2
(P4)
OFF
ON
Hysteresis (variable)
Pressure low
Pressure high
Vacuum low
Vacuum high
P1
(P3)
Hysteresis (fixed)
OFF
ON
P2
(P4)
Hysteresis (fixed)
(
Hysteresis
= 3 digits fixed)
Pressure low
Pressure high
Vacuum low
Vacuum high
1. Set value input mode
Press the "SET" button to display "P1-50(*1)".
The output OUT1 (1) set value input mode is selected.
(*1: If the set value of P1 is -50.)
2. OUT1 (1) set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
3.
Press the "SET" button to save the set value and select the output
OUT1 (2) set value input mode.
The set value of P2 is displayed.
4. OUT1 (2) set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
5.
Press the "SET" button to save the set value and select the failure
predictive pressure set value input mode.
The failure predictive set value is displayed.
6. Failure predictive pressure set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
7.
Press the "SET" button to save the set value and select the failure
predictive count set value input mode.
The failure predictive count set value is displayed.
8. Failure predictive count set value input
Pressing the
button increases the set value.
Pressing the
button decreases the set value.
9.
Press the “SET” button to save the set value and exit this mode.
1-output Type With the Failure Predictive Function
Note)
P1: Set value for OUT1 (1) P2: Set value for OUT1 (2)
P3: Set value for failure predictive pressure
P4: Set value for failure predictive count
The failure predictive detection counter is incremented when the
switch is turned on then is turned off, without the pressure (exceed-
ing P1) not reaching the failure predictive pressure (P3).
The failure predictive detection output is energized when the set fail-
ure predictive counter (EC) is incremented consecutively. When the
switch is turned ON and the pressure (exceeding P1) exceeds the
failure predictive pressure (P3), the failure predictive counter is reset.
(This example shows a case in the hysteresis mode.)
Failure Predictive Function
Degree of vacuum
High
Low
P3
P1
P2
Set pressure
Set pressure
Failure predictive count: Normal
Once
Twice
3 times
8. Outline dimensions (mm)
Dimensions
30
2
0
.5
01:R1/8
T1:NPTF1/8
L
e
a
d
w
ir
e
le
n
g
th
6
0
0
3
9
8
.5
3
.5
(3
0
0
0
)
13
M5 0.8
2- 3.5 mounting hole
1
0
30
2
0
.5
L
e
a
d
w
ir
e
le
n
g
th
6
0
0
4
0
.5
8
.5
3
.5
(3
0
0
0
)
13
01:R1/8
T1:NPTF1/8
M5 0.8
2- 3.5 mounting hole
Grommet Type
Connector Type
SE3-
-
C
I
Z
01
T1
SE3-
-
I
Z
01
T1
Single-installation Type
7
PV
2
0
.5
PV
50.6
Vacuum
supply
port
Vacuum
supply
port
Vacuum pad
connection
port
Air pressure circuit diagram
L
e
a
d
w
ir
e
le
n
g
th
6
0
0
5
1
.5
10
22.6
2- 3.5 mounting hole
(3
0
0
0
)
6
.5
1
8
5
8
8
10.25
0X:M5 x 0.8
0XY:M6 x 1.0
Vacuum supply port
0X:M5 x 0.8
0XY:M6 x 1.0
Vacuum supply port
0X:M5 x 0.8
0XY:M6 x 1.0
Vacuum pad
connection port
0X:M5 x 0.8
0XY:M6 x 1.0
Vacuum pad
connection port
5.75
1
0
.2
7
2
0
.5
50.6
5
0
1
0
10
22.6
6
.5
1
8
5
8
Vacuum pad
connection
port
Air pressure circuit diagram
5.75
1
0
.2
10.25
8
L
e
a
d
w
ir
e
le
n
g
th
6
0
0
2- 3.5 mounting hole
(3
0
0
0
)
Grommet Type
Connector Type
ZSE3- 0X
-
ZSE3- 0X
-
C
Peak Hold Mode
Pressing the
(UP) button when pressure is dis-
played enables the upper limit peak value (value with a high degree
of vacuum) to be held. In this case, "H" is displayed on the LCD. To
reset holding, press the
(UP) button again.
Bottom Hold Mode
Pressing the
(DOWN) button when pressure is
displayed enables the lower limit peak value (value with low vacu-
um) to be held. In this case "d" is displayed on the LCD. To return
holding, press the
(DOWN) button again.
Reset Function
Pressing the RESET button causes the following:
1) Measurement mode
•
Zero adjustment
•
Clearing the peak hold mode or bottom hold mode
•
Clearing the failure predictive function internal counter
•
Resetting the failure predictive output
2) Upon error occurrence
•
The data set in the setting mode is retained as is and the state
when the power supply was turned on is restored (System reset
is triggered).
•
In case of a data error, the setting mode is selected. When you
finish setting, the state when the power supply was turned on is
restored (System reset is triggered).
Note) In the set value input mode, the reset function does not work.
6. Other Functions
7. Names / Functions of Individual Parts
RESET button
SET button
(DOWN) button
(UP) button
LED
LCD
Main Unit
RESET button
: Resets at error occurrence and for 0 clear of
the display.
SET button
: Switches the setting mode and enters the set
value.
LCD
: Displays pressure value, setting mode, and
error code.
LED
: The green LED lights up when output OUT1
is ON. The red LED lights up when output
OUT2 is ON. When both OUT1 and OUT2 are
ON, both the green and red LEDs light up.
Upon error occurrence, the red LED flashes.
(UP) button
: Switches to the peak display mode and
increases the ON/OFF set value.
(DOWN) button
: Switches to the bottom display mode and
decreases the ON/OFF set value.