Details of Communication Format
HEC
4-5
4-6 Reading temperature average data
Note) For this product, the average temperature data, Tm, indicates the temperature of the external
sensor. Refer to “4-4 Reading external sensor” for the format.
4-7
Offset setting (without writing EEPROM)
Set offset of each Thermo-con from host. In this command, offset data loses effect with furning power to
OFF because it isn’t memorized into EEPROM.
It can be set at a minimum of every 0.01 deg.C.
Host
→
STX 36H
Offset
ETX
Check sum
CR
Symbol
10
0
10
-1
10
-2
(
H
)
(
L
)
←
Thermo-con
ACK
CR
Note)As for symbols, minus is indicated by “-“(=2DH) and plus is indicated by “0”(=30H).
<Specifying unit>
Host
→
SOH
UT STX 36H
Offset
ETX
Check sum
CR
Symbol
10
0
10
-1
10
-2
(
H
)
(
L
)
←
Thermo-con
ACK
UT CR
Note)As for symbols, minus is indicated by “-“(=2DH) and plus is indicated by “0”(=30H).
4-8 Reading offset
As the minimum unit is 0.01 deg.C, the value in the hundredths place can be read.
Host
→
ENQ
36H
Check sum
CR
(
H
)
(
L
)
←
Thermo-con
STX 36H
Offset
ETX Check sum
CR
Symbol
10
0
10
-1
10
-2
(
H
)
(
L
)
Host
→
ACK
CR
(Possible to omit)
Note)
As for symbols, minus is indicated by “-“(=2DH) and plus is indicated by “0”(=30H).