Figure 36: The user can customise the historical graphs using the “Options” button and selecting which
sensors to display and which to hide.
7.3
Auto Abort
The Auto Abort function allows users to set limits for toxic gases entering the analyser.
This helps to prevent damage to the equipment if excessive amounts of harmful gases such
as SO
2
, H
2
S, CO and HF enter the pipework. With the function set and enabled as soon as
any toxic gas level rises beyond the limit set, the test will abort, and the gas will be returned
to the compartment if auto-pumpback is enabled. To enter this function press “Menu”
followed by “Auto Abort” to see the following screen:
Figure 37: Auto Abort function. Depending on what toxic gas sensors are fitted a list of gases will be present
here.