PTE-50-CET
Sheet 25
Figure 10:
Channels Parallel Connection
Remember that the current outputs, to be added correctly, must have the same phase angle, for
instance:
We can select the same or different levels in the outputs but, to achieve the maximum power, we
recommend to use them with the same levels (for instance, if we want to inject 60 A, we would set 20
A in each channel).
To use the three power outputs as one, we will select the
Master-Slave-Slave configuration
, and then
we just have to turn the channel 1 output on.
Note: Of course, we can connect only two channels. It will depend on the power output required.
Channel 1
Channel 2
Channel 3
To
−
current
input in relay
To + current
input in relay
CH1
CH2
CH3
0.0
0.0
0.0