SMC Model 1026 Engineering Specification           

 

Rev. 100812              

 

 

Page 1 of 12                                                                                                                               

Schlyer Machine

 

Division of Audubon Machinery Corporation

 

814 Wurlitzer Drive 

North Tonawanda, NY 14120 

Toll Free (866) 867-7660 

fax: (716) 696-3174

 

http://www.schlyermachine.com 

 

Model CBW1026 Cage and Bottle Washer  

General Specifications 

 

 

1.0 

General Description 

 
 

The  Model  CBW1026  Cage  and  Rack  Washer  is  a  high  capacity  spray  cabinet  washer 

capable  of  processing  any  items  which  can  be  placed  inside  the  washing  compartment.  It  is 
specifically designed to clean and sanitize animal cages, bottles, pans, and other items ancillary to 
the care of laboratory animals. Several sizes are available, as well as many optional features. The 
Model  CBW1026  incorporates  a  large  number  of  innovative  and  unique  features  to  ensure 
thorough, efficient cleaning, as well as many features unique to our design. 
 

2.0 

Dimensions 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

       

(W”X H” X L”)            

   

(W”X H” X L”) 

MODEL 

DOOR 

CONFIGURATION 

CHAMBER SIZE* 

EXTERNAL 
OVERALL SIZE 

1026 

VERTICAL LIFT 

48" X 32" X 34" 

72" X 78" X 38-1/2" 

1026L 

VERTICAL LIFT 

48" X 32" X 48" 

72" X 78" X 52-1/2" 

1026LS 

LOAD SHELF 

48" X 32" X 34" 

72" X 78" X 38-1/2" 

1026LLS 

LOAD SHELF

 

48" X 32" X 48" 

72" X 78" X 52-1/2" 

1026C

 

CABINET 

48" X 32" X 34" 

72" X 78" X 38-1/2" 

1026LC

 

CABINET 

48" X 32" X 48" 

72" X 78" X 52-1/2" 

*Note: Dimensions are maximum chamber opening sizes. Actual load size must be smaller to 
allow for clearances.  Custom chamber sizes can be provided as required. 
 

3.0 

Reciprocating Spray Headers

  

 
3.1 

Vertical and horizontal headers, surrounding the load, shall be supplied and equipped with 

machined jets. Jets shall deliver an elliptical spray pattern for optimal impact and coverage to the 
load. No rotating spray arms shall be utilized. 
 
3.2 

Headers  shall  be  interconnected  and  mounted  to  a  carriage  having  non-lube  wheels 

constructed from single pieces of Delrin® homopolymer.  
 

 

Содержание CBW1026

Страница 1: ...ing as well as many features unique to our design 2 0 Dimensions W X H X L W X H X L MODEL DOOR CONFIGURATION CHAMBER SIZE EXTERNAL OVERALL SIZE 1026 VERTICAL LIFT 48 X 32 X 34 72 X 78 X 38 1 2 1026L...

Страница 2: ...viously entrained debris shall not be acceptable 4 5 Strainer shall be easily removable with no tools via a sanitary style clamp System shall be completely self regulating and require routine inspecti...

Страница 3: ...al 6 5 2 Door Open Lack of positive door closure shall cause the machine to enter Pause mode from where the current cycle can be recovered Fault message shall indicate which door is at fault if the ma...

Страница 4: ...shall be programmable from the Color Touch Screen PIN screens shall be available at the discretion of the supervisor to lock out access to the cycle phase programming functions All timers in the prog...

Страница 5: ...he last rinse shall be used to remove heavy soil flushing any easily removed matter to drain in order to ensure that the detergent solution stays as clean as possible Treatment shall be under pressure...

Страница 6: ...engineered part and sub assembly shall be scrutinized and all specific design decisions shall be made in the light of these basic criteria 30 1 3 All purchased components shall be un modified off of...

Страница 7: ...ets shall be elliptical spray jets equal to Spraying Systems Co 1 4USS8030 No proprietary jets shall be acceptable 30 10 Washer programmable control shall provided by a modular industrial type program...

Страница 8: ...provided with a door at each end for operation within a clean dirty room environment 40 5 Door Interlocks Doors shall be provided with pneumatically operated door locks arranged such that the two door...

Страница 9: ...de for Direct Factory Accessible Programming This option shall allow trained SMC technicians to directly access the programming of the machine from a remote location for changes or diagnostic purposes...

Страница 10: ...be chosen when available heights will not accommodate the standard doors in the open position 40 22 Load Shelf Type Door s Load Shelf Type door s shall be provided in order to match owner s existing...

Страница 11: ...ry shall be consulted for steam pressures below 30 psi dynamic 60 1 4 Hot water temperatures of less than 180 degrees F may impact treatment cycle times Factory shall be consulted for recommendations...

Страница 12: ...BW1026 1026LS 1026C CBW1026L 1026LLS 1026LC Shipping Wgt 2100 2400 Dynamic Wgt as Installed 1800 2000 Cage Capacity Std Mouse 32 48 Std Rat 12 16 Sump Capacity 25 Gals 25 Gals Heat Radiation typ door...

Отзывы: