ENGLI
SH
Temperature too low
Display:
„Lt“ & „Thermometer symbol“
The operating temperature is too low (below
the required temperature range from 10°C -
40°C (50°F - 104°F)). The error is a warning
that a continuation under these conditions
might lead to wrong blood glucose readings.
Solution:
Take the meter to a location with
appropriate working temperature and wait for
the next measurement until the error does not
occur again.
Test result is higher than 630 mg/dL (35.0
mmol/L)
Display:
„HI“
Solution:
Test again. If the result is still too
high, please contact your doctor immedia
-
tely.
Test result is under 20 mg/dL (1.1 mmol/L)
Display:
„LO“
Solution:
Test again. If the result is still too
low, please contact your doctor immediately.