54
Memory error
Display:
„005 Error“
Solution:
Replace batteries first. If “Err”still
occurs, please contact your local supplier.
Test strip already used or wet
Display:
„Err“ & „Test strip symbol“
Solution:
Use a new test strip.
Temperature too high
Display:
„Ht“ & „Thermometer symbol“
The operating temperature is too high (above
the required temperature range from 10°C -
40°C (50°F - 104°F)). The error is a warning
that a continuation under these conditions
might lead to wrong blood glucose readings.
Solution:
Take the meter to a location with
appropriate working temperature and wait
for the next measurement until the error does
not occur again.