TEST PROCEDURE HRD TEST

1.  Turn on machine.

2.  Keep PLT/HRD switch on “HRD”.

3.  Keep SET/RUN switch on “SET”.

4.  Set required Test Current (Amperes).

5.  Set Cutoff Voltage for test.

6.  For “HRD” test no need to set timer.

7.  Keep SET/RUN switch on RUN mode.

8.  Connect Battery Under Test.

9.  Machine will show test current,

battery volatge and zero timer.

CURRENT

VOLTAGE

TIMER

CURRENT

VOLTAGE

TIMER

CURRENT

VOLTAGE

TIMER

CURRENT

VOLTAGE

TIMER

10. Now Press START button or

foot switch (if provided) to start test.

11. Machine will put a load of a shown

test current and timer will start.

12. During test machine shows Green

and Red indication alternately.

CURRENT

VOLTAGE

TIMER

CURRENT

VOLTAGE

TIMER

CURRENT

VOLTAGE

TIMER

13. Test will end when battery voltage

drops below Cutoff.

15. After test end, LED blinking will stop,

and the time value is final HRD test timing.

CURRENT

VOLTAGE

TIMER

14.If HRD test continues for duration greater

than 59:59 minutes, HH:MM led is ON
and timer will show time in HH:MM.

CURRENT

VOLTAGE

TIMER

HH:MM led is ON

Time is 1 hour 36 Min

CURRENT

VOLTAGE

TIMER

For HRD test, timer setting is not considered
Timer will always start counting up from Zero.