A
Adjust control stop to meet the flow rate required for proper
cleansing of the fixture. Open control stop COUNTERCLOCKWISE
one (1) FULL turn from the closed position.
B
Activate flushometer by placing hand in front of sensor lens for ten
(10) seconds and then moving it away.
C
Adjust control stop after each flush until the rate of flow delivered
properly cleanses the fixture.
All Sloan Flushometers are engineered for quiet operation.
Excessive water flow creates noise, while too little water
flow may not satisfy the needs of the fixture. Proper
adjustment is made when plumbing fixture is cleansed
after each flush without splashing water out from the lip
AND a quiet flushing cycle is achieved.
Never open Control Stop to where the flow from the
valve exceeds the flow capability of the fixture. In the event
of a valve failure, the fixture must be able to accommodate
a continuous flow from the valve.
!!! IMPORTANT !!!
A
When an object is detected, a slowly flashing red light will appear
in the sensor window. After approximately eight (8) to sixteen (16)
seconds (closet) / eight (8) seconds (urinal), the light will flash
rapidly indicating sensor is armed and ready to activate solenoid
when the object leaves the detection area. The solenoid will be
activated within 0.5 seconds (for urinals) or three (3) seconds (for
water closets) after non-detection.
A
Turn power ON. The self adaptive sensor automatically adapts to the surrounding environment when 24 volt supply is activated. No manual
adjustments are required.
NOTE: IT IS RECOMMENDED THAT ALL ELECTRONIC CONNECTIONS BE TESTED WITH THE WATER SUPPLY OFF.
B
Start-up mode will take approximately one (1) minute to complete its cycle and is important that no non-permanent target is present at this time.
A continuous red light visible in sensor window indicates sensor is in the start-up mode. If the red light is flashing, this indicates that the sensor is
picking up a target. Unless this target is a permanent fixture in the sensor’s environment (i.e., a wall or stall door), it must be removed from the view
of the sensor. If this target is permanent, the sensor will adapt itself around this target. In this case, disconnect the 24 volt power supply for twenty
(20) seconds or more. Reconnect the 24 volt power supply at the transformer or fuse box. When the start-up cycle is complete there will be no light
visible in the sensor window.
NOTE: IF 24 VOLT POWER SUPPLY IS EVER INTERRUPTED FOR LONGER THAN TWENTY (20) SECONDS, THE START-UP MODE
AUTOMATICALLY BEGINS WHEN POWER IS RESTORED.
C
Incorrect wiring or a short in the 24 volt power supply is indicated by a continuous warning signal seen in the sensor window. The visible red light
flashes an “SOS” signal: three (3) short flashes, three (3) long flashes, three (3) short flashes and continually repeats this signal.
11 - DETECTION/ACTIVATION
13 - TURN WATER ON AND ADJUST CONTROL STOP
6
10 - POWER AND START-UP MODE
OPERATION
1. A continuous, invisible
light beam is emitted from
the sensor.
2. When a user enters the
beam’s effective range, for
water closets 22” - 42”
(559 mm - 1067 mm)
and for urinals 15” - 30”
(381 mm - 762 mm), the
beam is reflected into the
sensors scanning window
and transformed into a
low voltage
electrical signal
that activates a sixteen (16)
second (closet) /
eight (8)
second (urinal) time delay
circuit. The time delay
circuit eliminates false
operation from passers-by
in the rest room. Once the
time delay is completed, the
output circuit is alerted and
continues in a “hold” mode
for as long as the user
remains within the effective
range of the sensor.
3. When the user steps away
from the sensor, the loss of
reflected light initiates an
electrical “one-time” signal
that energizes the solenoid
operator, and activates the
flushometer to flush the
fixture. This occurs on the
water closet approximately
three (3) seconds after
indication. This delay is
built into the sensor to help
prevent false flushing due
to movement by the user.
The circuit for both water
closets and urinals then
automatically resets and is
ready for the next user.