Description of your unit’s symbols:
There are a number of technical markings on your unit. These can be
explained as follows:
The unit and belt are manufactured for Bio-Medical Research Ltd, Parkmore
Business Park West, Galway, Ireland.
The unit requires 1 x 3.6 volt (NiMH) DC battery pack DC is indicated by
the symbol:
The Output Frequency indicates the number of pulses per second transmit-
ted by the unit. This is measured in hertz, which is indicated by “Hz”.
Output (RMSA) means the maximum output root mean square current for
each channel.
Output (RMSV) means the maximum output root mean square voltage for
each channel.
This symbol means “Attention, consult the accompanying documents”.
This symbol means type BF applied parts
.
This symbol on your unit is to indicate conformity to the require-
ments of the Medical Device Directive (93/ 42/EEC). 0366 is the
number of the notified body (VDE).
SN stands for ‘serial number’. Inside the box, on the back of the unit is the
serial number specific to this unit. The letter preceding the number indicates
the year of manufacture, where “L” denotes 2006, “M” denotes 2007 etc.
The belt’s batch number is represented on the belt packaging, by the num-
ber corresponding with the LOT symbol.
Rated Outputs – Voltage/Currents: E:70 and X:70
Parameter
500Ω
1KΩ
1K5Ω
Output RMSV
7.35V
13.1V
15.5V
Output RMSA
14.7mA
13.1mA
10.3mA
Output Frequency
50-70Hz
50-70Hz
50-70Hz
DC Component: Approx.
0 C
0 C
0 C
Positive Pulse Width:
200-315µs
200-315µs
200-315µs
Negative Pulse Width:
200-315µs
200-315µs
200-315µs
Interphase Interval:
100 µs
100 µs
100 µs
Attention:
This unit can deliver currents in excess of 2mA/cm
2
if used with incorrect elec-
trodes.
At the end of the product lifecycle, do not throw this product into the
normal household garbage, but bring it to a collection point for the
recycling of electronic equipment.
Some product materials can be re-used if you bring them to a recycling
point. By re-using some parts or raw materials from used products you
make an important contribution to the protection of the environment.
Please contact your local authorities if you need more information
about collection points in your area.
Waste Electrical and Electronic Equipment can have potentially harmful
effects on the environment. Incorrect disposal can cause harmful toxins
to build up in the air, water and soil and can be harmful to human
health.
Belt Materials:
Outer material: 100% Nylon, Binding: 82% Nylon, 18% Elastane, Hook &
Loop: 100% Nylon, Foam: 100% Polyurethane
!
0366
14
technical specifications