Warning: Ensure that the detergent solution or water does not enter the unit or
the exhalation block gas ports on the side of the machine.
15.4 Disinfection method
Note: Alcohols such as 70% isopropanol have a good activity against bacteria
and viruses. They should only be used after all visible surface dirt has been
removed from the area to be disinfected.
Ventilator.
For disinfection use two clean, disposable, absorbent, non-shedding
cloths. Wipe clean with the first cloth using Alcohol (70% isopropanol). Wipe dry with
the remaining cloth.
Exhalation block.
The exhalation block can be immersed in Alcohol (70%
isopropanol). The exhalation block must be allowed to dry thoroughly before
sterilization.
15.5 Sterilization method
The silencer SLE part Nº N2186 and exhalation block SLE part Nº N2190 must be
sterilized between use on patients. The ventilator cannot be sterilized.
The exhalation block must be cleaned as an essential prerequisite to sterilization.
Autoclave with pure dry saturated steam at:
134ºC (277ºF) (Allowable variation of temperature of +3ºC) at 220kPa (32psi) with a
minimum holding time of 3 minutes
or
121ºC (248ºF) (Allowable variation of temperature of +3ºC) at 96kPa (14.1psi) with a
minimum holding time of 15 minutes.
There is no limit on number of autoclave cycles for the exhalation block.
The silencer can be autoclaved up to 20 times. The body of the silencer should be
marked after each autoclave cycle with a high temperature, water proof, permanent
maker to indicate number of sterilization cycles completed.