22
Resetting Pump to Manufacturer Default
Settings
1. Press the security code
and then press the CAL
bu
tt
on. The pump software version number will display
brie
fl
y.
2. Press
repeatedly to scroll through until the LED
displays dEFt.
3. Press [
]. The word no will appear on the display.
To reset to factory default se
tt
ings:
Press
. YES
will appear on the display. Press [
].
To retain existing se
tt
ings:
Press
to scroll to
no and press [
].
The number 30 will appear brie
fl
y, then the display will return
to run time presets.
Digital display:
Yes, reset to pump default
Digital display:
Pump default
Digital display:
No, do not reset to pump
default
Enabling/Disabling Alarm and Fault Features
1. Press the security code
to enter setup mode.
2. Press CAL. The pump software version number will display brie
fl
y.
3. Press
or
to turn feature on or o
ff
(see below)
. Press
to advance to next
feature.
4. Press [
] to exit feature setup. Press [
] again to return to run time
presets.
Feature ON
OFF
Button Beeper: On/Off
Alarm Beeper: On/Off
Flow Fault: On/Off
Flow Fault Restart: On/Off
Digital display:
Button beep on
Digital display:
Button beep off
Digital display:
Alarm beep on
Digital display:
Alarm beep off
Digital display:
Flow fault on
Digital display:
Flow fault off
Digital display:
Flow restart on
Digital display:
Flow restart off