10
Leland Legacy Quick Guide
Terms »
Star button
• Scrolls through run time data and Setup options
Up and down arrow buttons
• Toggle between display choices and increase or decrease sampling parameters in Setup
Button sequence
= press buttons individually
[
] = press simultaneously
= security code, always press in sequence
Security code
• Prevents unauthorized changes to the pump’s sampling program
Programming Sequences »
• To activate pump (e.g., to change pump from Sleep to Hold):
Press any button.
• To change pump from Hold to Run or Run to Hold:
Press [
].
• To reset accumulated data:
Press [
], then
. Press
un
til
CLr
di
splays then press [
]; press
until
End
displays
then press [
].
• To set pump
fl
ow rate:
Press [
], then
. Flow rate and SET
fl
ash. Press
or
to change
fl
ow rate. Press
until
End
appears then press [
] to save setting and place pump in Hold.
• To calibrate
fl
ow rate with standard calibrator:
Press [
], then
. Flow rate and SET
fl
ash. Press
or
to change
fl
ow rate. Press
once.
ADJ
displays. Press
or
until desired
fl
ow rate is indicated on calibrator. When
fi
nished, press
until
End
displays then press [
] to save new setting and place pump in
Hold. For CalChek Calibration, see
operating instructions.
• To change temperature scale from F to C or C to F:
Press [
], then
. Press
until temperature displays. Press
or
to switch units; press
until
End
displays then press [
] to save new setting.
• To change atmospheric pressure scale (mm, mb, In):
Press [
], then
. Press
until pressure displays then press
or
to switch units; press
until
End
displays then press [
] to save new setting.
• To change time scale (12 Hr/24 Hr/Dela):
Press [
], then
. Press
until 12 Hr, 24 Hr, or Dela displays then press
or
to switch
units; press
until
End
displays then press [
] to save new setting.
To set delayed start (Dela), see
operating instructions.
• To change clock:
Press [
], then
. Press
until clock displays then press
or
to change
fl
ashing hour; press
to move to minutes and
or
to change setting. Press
until
End
displays then press [
] to save
new setting.
• To change the sampling time function:
Press [
], then
. Press
until
ST L/min
displays then press
to change
fl
ashing digit; press
until
End
displays then press [
] to save new setting. To delete, follow above steps and press
until
0 appears. Exit Setup.
Note: When in Setup, choosing Esc instead of End will exit Setup without saving new settings.