Sampling

SKC tests indicate that a particulate mass of up to 6.8 mg on the four impaction 
substrates would not affect PPI performance. This amount is equivalent to sampling 
for 6 hours at 4 L/min in environments where respirable mass concentration is 
4.76 mg/m

and equals 50% of total dust. However, labs have reported to SKC 

that they prefer no more than 2 mg on the fi lter for analytical reasons. Therefore, 
SKC recommends that you work with your lab to determine optimum sample times 
for your unique sampling conditions.

1.  As per good industrial hygiene practice, replace 

representative sampler used for calibration with a 
new, unused loaded sampler. 

See Media and Sampler 

Preparation

.

2.  Record sample start time on label.

3.  Clip Disposable PPI onto a worker’s collar or lapel in 

the breathing zone or in the area to be sampled.

4. 

Clip sample pump at the worker’s waist or close to the 
Disposable PPI.

5. Use 

fl

 exible tubing to a

tt

 ach the Disposable PPI 

exhaust to the inlet of the sample pump.

6. 

Turn on pump and record pertinent sample data. 

7.  After the desired sample time has elapsed, turn 

o

   the pump and unclip sampler from sampling 

location.

8.  Apply Post-use label to inlet plate to seal the 

sampler.

9.  Record sample stop time on label on the bo

tt

 om 

of sampler.

10.  Disconnect sampler from pump and reinstall 

protective end cap on the exhaust.

11.  Reinstate calibration train with representative Disposable PPI and verify 

fl

 ow 

rate. 

See Calibration

.

3

6

-

8

10

10

Содержание Disposable Parallel Particle Impactor

Страница 1: ... respirable or thoracic and 4 or 8 L min respirable Sample Pump Universal or AirChek Series for 2 and 4 L min Leland Legacy for 8 L min Sample Time Dependent on method used Note SKC tests indicate that a particulate mass of up to 6 8 mg on the four impaction substrates would not affect PPI performance This amount is equivalent to sampling for 6 hours at 4 L min in environments where respirable mass...

Страница 2: ...the inlet plate Particles larger than each impactor s 50 cut point are scrubbed and retained by impaction onto the porous oiled impaction substrate contained in each impactor Smaller particles continue to the standard 37 mm collection filter for analysis PPI Performance Protective end cap Base plate Inlet plate contains 4 impac tors with unique cut points preloaded with 4 impaction substrates and ...

Страница 3: ...ps insertfirsta37 mmsupportpad and then a conditioned and weighed 37 mm collection filter if using gravimetric analysis into the base plate 3 Lay base plate on a flat surface align inlet plate with base and press down firmly to achieve an even seal 4 Write sample ID on sampling label Adhere sampling label to bottom of base plate 5 Remove protective end cap from exhaust Technical Tidbits Use forcep...

Страница 4: ...onnect the exhaust of the Disposable PPI to the inlet of a sample pump 5 Use flexible tubing to connect the inlet of the calibration adapter to the suction port outlet of a calibrator 6 Calibrate to 2 L min for the 2 L min respirable and thoracic model PPIs 4 L min for the 4 L min respirable model or 8 L min for the 8 L min respirable model Follow the calibration instructions in the pump and calib...

Страница 5: ...sed loaded sampler See Media and Sampler Preparation 2 Record sample start time on label 3 Clip Disposable PPI onto a worker s collar or lapel in the breathing zone or in the area to be sampled 4 Clip sample pump at the worker s waist or close to the Disposable PPI 5 Use flexible tubing to attach the Disposable PPI exhaust to the inlet of the sample pump 6 Turn on pump and record pertinent sample ...

Страница 6: ...ine Go to http dx doi org enter doi 10 1088 1742 6596 151 1 012060 Trakumas S High flow Personal Respirator Dust Sampler for Increased Sensitivity Poster 261 AIHce 2010 Denver CO Trakumas S High flow Personal Sampler to Monitor Exposure to Respirable Crystalline Silica at New Lower TLV IOHA 2010 8th Conference Book of Abstracts Rome p 59 Trakumas S Salter E High Flow Personal Sampler to Monitor Ex...

Страница 7: ...pore size pk 100 225 5 37 PTFE Filters 37 mm 2 0 μm pore size with laminated PTFE support for metalworking fluids NIOSH 5524 pk 50 Contact SKC MCE Filters 37 mm 0 8 μm pore size pk 100 225 5 Filter Supports required for User loaded Disposable PPI sampling Select either cellulose or stainless steel Support Pads cellulose 37 mm pk 100 225 27 Support Pads stainless steel 37 mm wide mesh ea 225 26 Acc...

Страница 8: ...8 Form 38055 Rev 1901 www skcinc com ...

Отзывы: