2
Principle of Operation
SKC Disposable PPI Samplers are impaction-based
fi
lter samplers that perform
precise size-selection for either thoracic or respirable dust, depending on the model.
PPI Samplers contain four small impactors in the inlet section of the device. Each
impactor features a unique 50% cut-point to target a speci
fi
c one-quarter segment of
the ISO/CEN curve resulting in a precise
fi
t along the entire curve. A sample pump
operating at 2, 4, or 8 L/min (2 L/min only for thoracic) pulls air through the inlet
nozzle of each impactor in the inlet plate. Particles larger than each impactor’s 50%
cut-point are scrubbed and retained by impaction onto the porous oiled impaction
substrate contained in each impactor. Smaller particles continue to the standard
37-mm collection
fi
lter for analysis.
PPI Performance
Protective
end cap
Base plate
Inlet plate (contains 4 impac-
tors with unique cut-points
preloaded with 4 impaction
substrates and sonically
welded)
Collar clip
Standard 37-mm collection
fi
lter (user installed)
37-mm support pad
(user installed)
Exhaust
to pump
Dust enters here
Collection ef
fi
ciency of the 2, 4, and 8 L/min
respirable PPI Samplers compared to the
ISO respirable curve
Collection ef
fi
ciency of the 2 L/min
thoracic PPI Sampler compared to the
ISO thoracic convention