
Application Scope, Operation Conditions, Safety and Cautions
1-6
device (e.g. a printer) from any plug before the system is turned off, otherwise it
may result in system damage or electric shock.
q) The system is not intended for use with a defibrillator.
r)
The system shall not be applied to cardiology directly.
s) Multiple devices, when interconnected, might result in accumulative leakage
current and dangers.
t) Make sure to use the special components provided by SIUI for system
component repair or replacement.
bed
patient
shield sheet
To the equipoten-
tial terminal of
main unit
Fig.1-1 Lay a shield sheet to eliminate the hum (AC noise) interference
1.6 Classification
a) Protection against electric shock
The system belongs to general equipment, Class I Type BF applied parts.
b) Protection against ingress of water
The system belongs to Class IPX0, and the probe head belongs to Class IPX7.
Содержание CTS-7700
Страница 1: ...DCY2 782 7700SS V5 0 B E Digital Ultrasound Imaging System OPERATION MANUAL ...
Страница 2: ......