
12
Measurement Diagram
Manual de Instruções
1
2
3
100%
0%
MIN. LEVEL /
DISPLAY
MAX. LEVEL /
DISPLAY
1 - Dead zone or instrument erasure, (when level is above the dead zone a
DEAD ZONE
error message will show in the display
indicating that above this range the instrument will not measure) see (page 23) technical specifications to know the blind area of
the instrument.
2- Maximum distance or measuring distance to be respected for each transmitter model (eg SLU 02 measures up to 2 meters /
SLU06 measures up to 6 meters / SLU10 measures up to 10 meters / SLU20 measures up to 20 meters)
3 - Measuring range (Measuring range to be set on the transmitter in the parameters
MIN. LEVEL
0% and
MAX. LEVEL
100%
(with reference to the face of the transducer) and
DISPLAY
parameters by entering the value of the process that you want to show
in the display
Note: The measuring range to be configured must not be less than dead zone (1) and greater than the maximum ultrasound
measurement distance (2).
Reference (face of the transducer)
Содержание SLU
Страница 11: ...11 Set up elements Status Signalization Installation Manual ...
Страница 13: ...13 Basic configuration Operation and Setting Sitron Installation Manual ...
Страница 14: ...14 Installation Manual ...
Страница 15: ...15 Installation Manual ...
Страница 20: ...20 Protocol HART SLU Protocol Modbus SLU M SMC 485 Installation Manual SLU XX SLU ...
Страница 22: ...22 Installation Manual SLU XX SLU Menu Structure ...