11.5 Check and adjust optical plummet
Adjustment is required to make the line of sight of
plummet telescope coincide with the vertical axis (as otherwise
the vertical axis will not be in the true vertical when the instru
-ment is optically plumbed.)
Check
optical
a. Coincide the center point with the center mark of optical
plummet telescope by adjusting optical plummet.
b. Revolve the instruement 180
or 200g around the vertical
axis and check the center mark. If the point is properly center
-ed in the centermark, adjustment is not required. Otherwise,
adjustment will be done in the following procedures.
Adjustment
a. Unscrew the adjustment section cover of the optical plummet
telescope eyepiece, by revolving it in the counter clock-wise
direction and take it off. this will expose four capstan adjust
-ment screws which should be adjusted with the accessory
adjusting pin to shift the center mark to the point. However,
correct only 1/2 of the displacement in this manner.
o
26
b. Next use the leveling screws and coincide the point and
center mark.
adjustment screw
optical plummet