can measure and adjust tilt compensator 0 position error. and
you can measure collimation error in your instrument so that
the instrument can correct subsequent single face observations.
The 0 index of the vertical circle of your instrument can be also
reset, and the index error that will affect the accuracy of vertical
angle measurement can be corrected.
18
operating display
【
R/L
】
SET F1
HA
R
0° 00
′
08
″
1. Press and
power, it will display
“
SETUP
” and “
SET
0”,
rotate the telescope,
the first line will display
“
SET F1
” and will glint.
11-03-20 15:48
if you want to exit at any time, you can press
.
【
】
【
【
】
】
2. Level the instrument and
make the plate at left,
collimate the target at
infinitude press ,
the first line will glint and
display “SET F2”.
OSET
OSET
【
】
OSET
3. Make the plate at right
and then collimate the
same target, press
the first line will glint
and display “SET”.
4. Press the instrument
perform the new data of the
vertical error, the telescope
axis error and the
compensator 0 error, and
return to the angle mode.
11-03-20 15:48
SET F2
HA
R
0° 00
′
08
″
SET
HA
R
179° 59
′
58
″
11-03-20 15:48
SFT
Note: After adjustment above finished, you should check again.