S.Cod.
d.C.sc.
Uo.sc.
D.Ca.s.
CAL.t.
= Calibration type.
Selectable as
S.inP
.
or
.
If
S.inp
.
is selected, input type is one the
four standard input types. If
U.inP
is
selected, input types can be modified.
See NOTE 1 for modification.
U.inP
d.Pnt.
= Decimal point.
Decimal point can be adjusted between
1. and 3. digits.
See NOTE 1 for programming.
s.Cod.
= Access code for calibration menu.
This parameter should be
222
.
See NOTE 1 for programming.
s.Cod.
= Access code for safety menu.
This parameter should be
333
.
See NOTE 1 for programming.
d.C.Sc.
=
ConF.
menu protection level
parameter.
nonE
= No menu is seen.
P. no
= Menu is seen but can not be programmed.
P.yES
.
= Menu is seen and programming is possible.
See NOTE 1 for programming.
Uo.Sc.
=
U.oPt.
menu protection level
parameter.
nonE
= No menu is seen.
P. no
= Menu is seen but can not be programmed.
P.yES
.
= Menu is seen and programming is possible.
See NOTE 1 for programming.
d.Ca.S.
=
d.CAL.
menu protection level
parameter.
nonE
= No menu is seen.
P. no
= Menu is seen but can not be programmed.
P.yES.
= Menu is seen and programming is possible.
See NOTE 1 for programming.
L.SCL.
= Lower limit for the scale.
It can be adjusted between -1999 and
(
H.SCL
.
-100).
See NOTE 1 for programming.
H.SCL.
= Upper limit for scale.
It can be adjusted between (
L.SCL.
+100)
and 4000.
See NOTE 1 for programming.
S.Cod.
a.CAL.
1.CA.
10.CA.
For adjusting a selected parameter first press and hold key. Then, by using keys adjustment
can be made.
Parameter adjustment method
CAL.
CAL.
C.end
C.end
s.err.
S.err.
C.err.
C.err.
The message on the left flashes approximately 5
seconds and calibration is completed.
The message on the left flashes approximately 5
seconds and calibration is completed.
If calibration is error free, the message
C.End
appears
. However,
if it is wrong, the message
S.Err
.
appears for 1 seconds and the program
is shifted to the next step.
for 1 seconds
If calibration is error free, the message
C.End
appears .
However,
if it is wrong, the message
C.Err.
appears for 1 seconds and the program
is shifted to the next step.
for 1 seconds
If the difference between the reference voltages or currents applied for the calibration of
H.inP.
and
L.inP.
is lower than one
half of the full scale, this error message appears on the display.
For example: Assume that the selected input type is 0-1V. In this case, if the difference
voltages
applied for calibration of
H.inP.
and
L.inP.
is lower than 0.5V, this error message appears.
between the reference
If the reference voltage or current applied to the input for calibration is too high or too low, this error message appears.
ERROR MESSAGES
CAL okey?
CAL okey?
CAL okey?
Evet
Evet
Hayýr
Hayýr
NOTE 2
NOTE 3
If first and then keys are pressed together.
programming mode is entered.
U.opt.
d.Cal.
SeCU.
d.pnt.
Cal.t.
l.sCl.
H.sCl.
l.inp.
H.inp.
i.typ.
= Input type.
Input type can be selected as 0-20mA,
4-20mA, 0-1V or 0-10V.
See NOTE 1 for programming.
dSP.C.
= Display configuration.
Selectable as
PrcS.
or
Pr.Un.
. If
PrcS.
is
selected, process value appears. If
Pr.Un.
is selected, process value and then
measurement unit are displayed 4 and 2
seconds successively.
See NOTE 1 for programming.
RAtE
= Sampling rate.
Measurement is performed at each 200ms.
However,
for
FASt
rate, each measurement is displayed.
for
SLo.1
rate, the average of 4 successive
measurements is displayed.
for
SLo.2
rate,
for
SLo.3
rate,
the average of 8 successive
measurements is displayed.
the average of 16 successive
measurements is displayed.
See NOTE 1 for programming.
Hold
= Display holding parameter.
Selecting
NonE
, this parameter becomes
inactive.
Selecting
Lo.
, always the minimum
measured value is displayed.
Hi.
,
Selecting
always the maximum
measured value is displayed.
CAL.t.=U.ýnP.
Evet
Hayýr
Unit
= Measurement unit.
A constant, a message etc. to be displayed
can be entered. If a decimal point is desired,
it should be included before entering the
character.
For including decimal point first, then,
keys are pressed and held together. And
then, by using key decimal point can be
adjusted.
d.Cnf.
i.typ.
rate
Hold
dsp.C.
Unit
At this state, the reference voltage or current
that corresponds to
L.SCL.
parameter is applied
to the input.
To initialize calibration, first then
keys are pressed together and held until ‘
'
message appears. See NOTE 2.
CAL
At this state, the reference voltage or current
that corresponds to
parameter is applied
to the input.
To initialize calibration, first then
keys are pressed together and held until ‘
'
message appears. See NOTE 2.
H.SCL.
CAL
A.CAL.
= Current calibration.
At this state, 20.000 mA current is applied to the
input of the device.
10.CA.
= 10V input calibration.
At this state, 10.000V is applied to the input of
the device.
1.CA.
= 1V input calibration.
At this state, 1.0000V is applied to the input of
the device.
are pressed together,
measurement unit appears. See
Unit
parameter for programming.
If first and then keys
are pressed together, the
maximum and the minimum
measurement values become
equal to the measured value at
that time and the message
.
Res.
appears on the display
Pressing this key the minimum
measured value up to that time
appears.
Pressing this key the maximum
measured value up to that time
appears.
The minimum
measured value
The maximum
measured value
Measurement unit
870
1453
Bar.
&
Programming mode
&
3/3
MAX.
UNIT
MAX.
UNIT
MAX.
UNIT
MAX.
UNIT
MAX.
UNIT
MAX.
UNIT
RESET
RESET
RESET
RESET
MAX.
UNIT
RESET
MAX.
UNIT
RESET
RESET
RESET
RESET
RESET
RESET
PRG
RESET
MAX.
UNIT
SHIFT
SET
MAX.
UNIT
MAX.
UNIT
MAX.
UNIT
MAX.
UNIT
MAX.
UNIT
RESET
RESET
RESET
RESET
RESET
RESET
MAX.
UNIT
SHIFT
MAX.
UNIT
SHIFT
SET
For initializing calibration, first then
keys are pressed together and held until ‘
'
message appears. See NOTE 3.
CAL
To initialize calibration, first then
keys are pressed together and held until ‘
'
message appears. See NOTE 3.
CAL
To initialize calibration, first then
keys are pressed together and held until ‘
'
message appears. See NOTE 3.
CAL
MAX.
UNIT
SHIFT
SET
MAX.
UNIT
SHIFT
SET
MAX.
UNIT
SHIFT
SET
SET
RESET
RESET
UNIT
MAX.
UNIT
MAX.
UNIT
SHIFT
SET
SHIFT
SET
MAX.
UNIT
MAX.
UNIT
PRG
MIN.
PRG
MIN.
RESET
PRG
MIN.
PRG
MIN.
PRG
MIN.
SHIFT
SET
SHIFT
SET
SHIFT
SET
SHIFT
SET
Run mode
MAX.
UNIT
RESET
PRG
MIN.
Measurement
value
1000
Res.
EI741A-E-05
IF first, and then keys
NOTE 1
If increment key is pressed and held 0.6 seconds, the value of the selected parameter changes rapidly. If
waited enough, the value increases 100 at each step. After 1 second following the release of the key, initial
condition is returned. The same procedure is valid for the decrement key.