Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
6 Maintenance and inspection
64 08 137 D3610
D3610.201.02.09.02 11.2017
103
6.2
Regular inspections
Inspections must be performed at scheduled intervals to protect the
health and safety of patients, users and other persons.
Before and during operation
The system owner or an appointed person must ensure that all
components such as cables, sensors and housing parts are in sound
condition.
Monthly
Once a month the system owner or an appointed person must:
● check the sensor cable thoroughly for wear and tear
● make sure the connector housing is fastened securely at the sensor
cable
Annually
The image quality must be assessed by the system owner or an
appointed person at regular intervals, at least once a year.
On digital sensors, the degree of postprocessing (brightness or contrast
adjustment) that is required in the image processing software (e.g.
SIDEXIS XG / SIDEXIS 4) to produce satisfactory results is used as an
assessment criterion.
If these assessment criteria are regarded as given irrespective of the
patient’s anatomy and/or possible sources of error such as patient
positioning, a service engineer should be called in immediately to rectify
any possible system faults.
CAUTION
Should you identify faulty components on your XIOS XG product,
please inform your appointed service engineer. In this case, the
device should no longer be used for patient exposures.
Observe any possible additional country-specific requirements.
Please also check that all labels on the underside of the charger are
undamaged and legible.
CAUTION
The WiFi interface and the charger must not be opened or repaired by
the user.
All parts of the device are maintenance-free. In case of
malfunctioning, please always contact your specialized dealer.