5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Vacuum system
Foaming cleaning agents
Preparation for cleaning
1.
Prepare 1 liter of cleaning and disinfectant solution in a separate
container according to the manufacturer's instructions and mix it
thoroughly.
2.
Remove the gold trap
(A)
.
3.
Clean the cuspidor bowl.
4.
Insert the cleaning adapter
(B)
as far as it will go.
5.
Plug the saliva ejector hose
(D)
onto the adapter
(C)
and insert the
adapter in the cleaning adapter
(B)
.
Cleaning procedure
1.
Pour 1 liter of cleaning and disinfectant solution into the cuspidor bowl.
2.
Remove suction hoses
D
and
E
from their holders and attach them to the
side of the cleaning adapter
(B)
as simultaneously as possible.
ª
The cleaning and disinfectant solution is aspirated.
3.
Wait for a while to let the cleaning and disinfectant solution react.
Observe the reaction time specified for the cleaning and disinfectant
solution by the manufacturer.
CAUTION: Domestic cleaning agents foam up.
Foaming cleaning agents often cause foam and water to be sucked into the
dry suction system. This can cause damage to the suction machine.
¾
Use only care and cleaning agents which have been approved by Sirona,
see "Care and cleaning agents" [
B
A
1LDesinf
.
E
D
C
B