8 Electrical connection
Installation Manual ORTHOPHOS XG 3D / Ceph
8.2 Connecting the line voltage
båÖäáëÜ
Power connection: XG 3 / 5 / 3D / 3D ready
1. Check the protective ground wires and the unit leakage current
according to IEC 62353 (see section Checking the protective ground
wires and Checking the device leakage current.)
2. Record the measured values in the document titled Inspection and
maintanance and safety-related checks (REF 63 03 601).
3. Now connect the power supply as shown.
8.2.2
Connecting the media converter
➢ Plug the connector of the media converter's power supply unit into the
electric outlet.
CAUTION
Incorrectly connected units can pose a risk to patients
Building installation with 3x1.5 mm² or 3x2.5 mm² (16 AWG or 14 AWG)
and a 16 A/20 A overcurrent circuit breaker.
➢ Connect only units which do not pose any risk to patients when the
automatic circuit breaker is triggered.
➢ Do not connect any EDP units.
➢ See also the documentation titled Installation Requirements (REF
63 03 551).