ULTRAPEEN
3. SYSTEM SET-UP
3. SYSTEM SET- UP
CAUTION:
Plugging the Ultrasonic Generator unit into a socket, which supplies improper
voltage and/or current rating, may cause the Generator to malfunction or create an
electric shock or fire hazard.
CAUTION:
The Generator Electrical Supply cord should not be plugged into a device (e.g.
“power strips”, “gang plugs”, etc.) other than an industrial grade wall socket or a standard
gang plug with an appropriate wire section and current rating. Any
other unauthorized
use could cause significant changes in voltage that could result in an electrical fault
indication.
3.1.
MAINS POWER CONNECTION
Verify that the Mains Power Switch “
Power”
on the front panel of the Generator is in the
OFF position.
Plug the female end of the supplied grounded line cord into the Mains Power Supply
connector in the rear of the Generator (see Figure 2). Plug the other end of the line cord
into a properly grounded three prong 220-240 VAC (single phase) or 110-120 VAC
grounded socket receptacle. UP-600 system with output power up to 500 Watts requires
a mains power circuit with a nominal
fuse
rating of
8.0
А
mps
for 110/120 VAC and
4 Amps for 220/240VAC
.
3.2.
TOOL POWER CONNECTION
The ultrasonic frequency voltage is supplied to the
Tool
via permanent connector in the left
lower corner on the Generator’s front panel (see Figure 1).