FOR Safe AQ Voice meter

1. If test results greater than 600 mg/dL

33.3mmol/L

, the meter will display 

“HI” and read“The result is beyond the range”with red light on. Retest your blood 

glucose immediately with a new test strip. If your reading is still “HI”, contact 

healthcare professional immediately.

2. If test results is lower than 20 mg/dL

1.1mmol/L

, the meter will display “LO” 

and read “The result is below the range”with red light on. Retest your blood 

glucose immediately with a new test strip. If your reading is still “LO”, contact 

healthcare professional immediately. 

3. If test result is >70mg/dL

3.9 mmol/L

and 

 126 mg/dL

(<

 7.0 mmol/L

the meter will display and read the test result with green light on. 

4. If test result is ≥ 126 mg/dL

≥7.0 mmol/L

and ≤ 250 mg/dL

≤13.9 mmol/L

the meter will display and read the test result with yellow light on. 

5. If test result is ≤ 70mg/dL

≤ 3.9 mmol/L

or >250mg/dL

> 13.9 mmol/L

the meter will display and read the test result with red light on. 

Note: Do not change your treatment only based on test result and light indications. 

Please consult healthcare professional for your treatment. 

V.About Your Safe AQ Smart

Safe AQ Air

Safe AQ Voice Meter

Test

 

strip

 

port

 

Strip

 

eject

 

b

utton

LCD

 

display

Main

 

button

Test

 

strip

 

port

 

Strip

 

eject

 

b

utton

LCD

 

display

Main

 

button

Battery Cover

Battery Cover

Safe AQ Smart Meter: 

Safe AQ Air Meter:

Содержание Safe AQ Smart

Страница 1: ...VPQ Shanghai International Holding Corp GmbH Europe Eiffestra e 80 20537 Hamburg Germany Tel 49 40 2513175 Fax 49 40 255726 E mail shholding hotmail com No 265 Guyuan Road Hi Tech Zone Changsha Hunan...

Страница 2: ...r manual contains important information about the meter and how it works Please read this user manual carefully before using your new meter or the protection provided by the meter may be impaired If y...

Страница 3: ...Voice only 10 VIII Pairing Connecting Meter with Smartphone For Safe AQ Air only 11 IX Control Solution Testing 12 X Blood Glucose Testing 16 XI Viewing test result 23 XII Memory 24 XIII Information f...

Страница 4: ...used for the diagnosis of or screening of diabetes The Safe AQ Smart Blood Glucose Monitoring System includes Safe AQ Smart blood glucose meter Safe AQ blood glucose test strip blood glucose control...

Страница 5: ...ter and lancing device are for single person use only Do not share them with anyone else even family members For professional use if the meter is being used on the second person the meter and lancing...

Страница 6: ...ance with your feeling retest your blood glucose If still out of the range please contact healthcare professional FOR Safe AQ Smart Safe AQ Air meter 1 If test results greater than 600 mg dL or 33 3mm...

Страница 7: ...result is 70mg dL 3 9 mmol L and 126 mg dL 7 0 mmol L the meter will display and read the test result with green light on 4 If test result is 126 mg dL 7 0 mmol L and 250 mg dL 13 9 mmol L the meter...

Страница 8: ...is preset as voice on Slide the battery cover out of the meter Remove the batteries Voice switch is inside the battery chamber Slide the voice switch to the on side voice turns on and screen will disp...

Страница 9: ...rent App may be little different while pairing meter with smartphone IX Control Solution Testing Why Perform a Control Solution Test Ensures that your meter and test strips are working properly Allows...

Страница 10: ...ly the Control Solution 1 Check the expiration date and discard dates on your control solution and Safe AQ test strip vials Do not use control solution or test strip which expired 2 Shake the control...

Страница 11: ...ake control solution well and retest with a new test strip Retest with a new test strip Change a new vial control solution and retest Change a new vial test strip and retest Equilibrate meter test str...

Страница 12: ...rage skin 5 deepest depth for thick or calloused skin STEP 7 Pull the spring cover back to arm the lancing STEP 8 Take a strip out of the strip vial Immediately close the vial lid Attention Finish tes...

Страница 13: ...y The temperature is below the range it means the environment temperature is below the range Please put the meter strip control solution at appropriate place 10 35 wait at least 30 minutes and test ag...

Страница 14: ...for Safe AQ Voice meter the meter displays and read the test result The result is xxx Milligram per deciliter or The result is xx point xxx millimol per liter STEP 13 Eject the strip Caution DO NOT s...

Страница 15: ...screen displays HI Use a new test strip to test again If still display HI please contact healthcare professional immediately If test result is below 20 mg dL 1 1mmol L the screen displays LO Use a new...

Страница 16: ...er be used on multiple patients there is risk of infection between both patients and healthcare professional Water or other disinfection solution which remains on skin may dilute blood and cause inacc...

Страница 17: ...9935582 or contact local dealers To disinfect the meter Please according to clinic institution disinfect method to disinfect the meter XVI Power Saving If there is no blood applied to the test strip w...

Страница 18: ...tion conditions Dimension Weight Power source Software Version A 20 55 84 60 26 mm 73 5g with battery 3V DC 50mA 2 AAA alkaline batteries 5s Memory 10 blood glucose test results Testing condition 10 3...

Страница 19: ...nd test levels specified by IEC 61326 2 6 2012 Do not use this instrument in close proximity to sources of strong electromagnet ic radiation as these may interfere with the proper operation Avoid dirt...

Страница 20: ...sufficient samples Use a new test strip to test again Use a new test strip to test again Use a new test strip to test again Call Customer Service at 86 731 89935581 86 731 89935582 or contact your loc...

Страница 21: ...rip exposed in air too long after opening the vial or package The end of test strip may press too firmly to your finger blood sample cannot be drawn into test strip reaction chamber The test strip pac...

Страница 22: ...st strip may be affected with damp Test result is HI or LO but no red light on Only for Safe AQ Voice Test result is 70mg dL 3 9 mmol L and 126 mg dL 7 0 mmol L but no green light on Only for Safe AQ...

Страница 23: ...away from sunlight Serial number Authorized representative in the European Community CE Marking and Notified Body Number Biological risks Symbol for the marking of electrical and electronics devices...

Страница 24: ...This action will avoid the warranty and cause the meter to display false results 4 If you meter does not work properly please fill in warranty card carefully Use 75 alcohol to clean your meter and sen...

Страница 25: ...lease do complete the warranty card and send it back to our agent or our factory we will inform you with information of the meter Thank you for supporting our product and us Thank you for choosing Saf...

Страница 26: ...45...

Отзывы: